Notes concerning the catchpot at the inlet (AK)
The catchpot at the inlet prevents droplets and particles from entering the pump.
- Lifetimes of diaphragms and valves are enhanced.
- Improves vacuum in case of condensation.
- The round bottom flasks are coated with a protective layer to prevent disintegration in case of break-
age or implosion.
• In case of condensation: Check liquid level in both catchpots during operation. Avoid overflowing of
the catchpots. Do not overfill the catchpots. Maximum liquid level approx. 80%, to avoid problems
when removing the catchpots.
• Check liquid level in both catchpots regularly and drain catchpots in time.
Removing the catchpot at the inlet:
Admit air or inert gas (via inlet of pumping unit) to atmospheric pressure.
Remove joint clip, remove catchpot and drain condensate.
Reassemble drained catchpots.
Important:
Comply with regulations when disposing solvents/condensates. Reuse if possible, purify if
contaminated.