background image

 page 5 of 20

Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully 

the validity of this document with respect to his product.  11/08/2015  //  999281

In case of devices which have been in contact in contact with biological substances 

of risk level 2 contact the VACUUBRAND service absolutely before shipping the 

device. These devices have to be completely disassembled and decontaminated by 

the user prior to shipment. 

Do not return devices which have been in contact with biological substances 

of risk level 3 or 4. 

These devices cannot be checked, maintained or repaired. The 

devices must not returned to VACUUBRAND due to a residual risk in spite of decon-

tamination. 

The same conditions apply to on-site work.

This manual describes the decontamination of the aspiration systems BVC and the 

handle VHC

pro

.

The applicability of the decontamination method and the selection of the ap-

propriate disinfectants have to be ensured by the user.

Ensure that work is done only by an appropriately trained and supervised personal. 

Obey local and national safety requirements. Ensure that the personal is familiar 

with the safety procedures which relate to the products processed by the pumping 

system. Ensure that the device is decontaminated before repair and that you take 

adequate precautions to protect people from the effects of hazardous substances if 

contamination has occurred.

+

 

Take adequate precautions to protect people from the effects of dangerous sub-
stances. Wear appropriate personal protective equipment (e. g., protective clothing, 
safety glasses, protective gloves) and perform work in a hood.

Prior to decontamination 

isolate equipment from mains. 

Allow sufficient cooling. 

Ensure that the device cannot be operated accidentally. 

Wait 5 seconds

 before starting work to allow the capacitors to discharge. 

 

Before starting check that the required parts (e. g. tools, disinfectants) are available 

and of the correct type.

Check the operating sequence mentally on feasibility, safety requirements and con-

sequences on safety and function of the equipment. 

Obey regulations when disposing of solvents and 

chemicals

.

In order to comply with law (occupational, health and safety regulations, safety at 

work law and regulations for environmental protection) vacuum pumps, components 

and measuring instruments returned to the manufacturer can be repaired only when 

certain procedures (see section 

”Repair - Maintenance - Return - Calibration”

are followed.

Scrapping and waste disposal

Dispose of the pump and any components removed from it safely in accordance with 

all local and national safety and environmental requirements.

Safety information!

NOTICE

IMPORTANT

Summary of Contents for BVC control

Page 1: ...nged It is strictly the users responsibility to check carefully the validity of this document with respect to his product 11 08 2015 999281 Technology for Vacuum Systems Decontamination of BVC basic B...

Page 2: ...d distributed completely and unchanged It is strictly the users responsibility to check carefully the validity of this document with respect to his product 11 08 2015 999281 After sales service Contac...

Page 3: ...ty to check carefully the validity of this document with respect to his product 11 08 2015 999281 Contents Safety information 5 Materials 6 Decontamination of the vacuum pump 8 Disassembling the BVC c...

Page 4: ...y the users responsibility to check carefully the validity of this document with respect to his product 11 08 2015 999281 WARNING indicates a hazardous situation which if not avoided could result in d...

Page 5: ...ng system Ensure that the device is decontaminated before repair and that you take adequate precautions to protect people from the effects of hazardous substances if contamination has occurred Take ad...

Page 6: ...FE PP PPS POM ETFE aluminium alloy Feet NBR BVC control G BVC professional G Support cover pump PBT Pump housing base plate aluminium alloy On off switch PC EPTR IEC plug PET Feet PU Screws stainless...

Page 7: ...ETFE Outlet ETFE Silencer silicone rubber Sensor PPS fuorosilicone Sensor tube PTFE PVC Collection bottle 2l Bottle borosilicate glass Cap insert PP Flat seal EPDM Hose nozzle at bottle head connecti...

Page 8: ...minated parts Adopt suitable measures to prevent the release of dangerous toxic explosive cor rosive noxious or polluting fluids vapors and gases and a possible contamination of personal and environme...

Page 9: ...aphragm key SW 46 order no 637777 Screw driver Torx TX20 Open ended wrench SW 19 Tweezers Connect person to ground before coming in contact with electronic parts IMPORTANT Disassembling the BVC contro...

Page 10: ...tly the users responsibility to check carefully the validity of this document with respect to his product 11 08 2015 999281 Unscrew hose nozzle or hose nozzle with coupling using an open ended wrench...

Page 11: ...ty of this document with respect to his product 11 08 2015 999281 View of the disassembled pump head parts ME 1C 1 Cover of housing cover 2 Housing cover 3 Housing cover insert 4 Valves 5 Head cover 6...

Page 12: ...m to its lowest oscillating position Unscrew di aphragm support disc with diaphragm and diaphragm clamping disc If the diaphragm is difficult to separate from the diaphragm support disc immerse assemb...

Page 13: ...ment with respect to his product 11 08 2015 999281 Lay down the BVC as shown in the figure Unscrew screws at the base plate Slide the base plate to the side Remove earth cables Remove hose from pressu...

Page 14: ...20 Documents are only to be used and distributed completely and unchanged It is strictly the users responsibility to check carefully the validity of this document with respect to his product 11 08 20...

Page 15: ...mpletely and unchanged It is strictly the users responsibility to check carefully the validity of this document with respect to his product 11 08 2015 999281 Disassembling the BVC basic Disassemble th...

Page 16: ...users responsibility to check carefully the validity of this document with respect to his product 11 08 2015 999281 Disassembling the VHCpro Turn cap 1 and remove 2 Pull adapter with hose out of the...

Page 17: ...pletely and unchanged It is strictly the users responsibility to check carefully the validity of this document with respect to his product 11 08 2015 999281 It is not possible to disassemble the VHCpr...

Page 18: ...no original components Drain the device completely of fluids and residues Decontaminate the device Close all openings airtight especially if using substances hazardous to health To expedite repair an...

Page 19: ...r service staff 11 If the paint is damaged we wish a repaint or a replacement of parts for reason of appearance repaint and replacement at customer s expense yes no 12 Legally binding declaration We a...

Page 20: ...btained under lab conditions to your specific operations depends on a number of factors beyond our control It is therefore strictly the users responsibility to very carefully check the application of...

Reviews: