Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully 

the validity of this document with respect to his product.  11/08/2015  //  999281

Every employer (user) is held responsible for the health and safety of his employ-

ees. This also applies to service personnel performing repair, maintenance, return 

or calibration.

The 

health and safety clearance form

 informs the contractor about any possible 

contamination of the device and forms the basis for the risk assessment. 

In case of devices which have been in contact with biological substances of 

risk level 2 contact the VACUUBRAND service absolutely before dispatching 

the device. These devices have to be completely disassembled and decon-

taminated by the user prior to shipment. Do not return devices which have 

been in contact biological substances of risk level 3 or 4. 

These devices cannot 

be checked, maintained or repaired. Also decontaminated devices must not returned 

to VACUUBRAND due to a residual risk.

The same conditions apply to on-site work.

No repair, maintenance, return or calibration is possible unless the correctly 

completed health and safety clearance form is returned. Devices sent are re-

jected if applicable. 

Send a completed copy of the 

health and safety clearance 

form

 to us in advance. The declaration must arrive before the equipment. Enclose a 

second completed copy with the product. 

Remove all components from the device that are no original VACUUBRAND compo-

nents. VACUUBRAND will not be responsible for lost or damaged components that 

are no original components. 

Drain the device completely of fluids and residues. Decontaminate the device.

 

Close all openings airtight especially if using substances hazardous to health. 

To expedite repair and to reduce costs, please enclose a detailed description of the 

problem and the product’s operating conditions with every product returned.

If you do not wish a repair on the basis of our 

quotation

, the device may be returned 

to you disassembled and at your expense.

In many cases, the components must be cleaned in the factory prior to repair.

For cleaning we use an environmentally friendly water based process. Unfortunately 

the combined attack of elevated temperature, cleaning agent, ultrasonic treatment 

and mechanical stress (from pressurised water) may result in damage to the paint. 

Please mark in the health and safety clearance form if you wish a repaint at your 

expense just in case such a damage should occur. We will also replace parts for cos-

metic reasons at your request and at your expense.

Before returning the device

Pack the device properly, if necessary, please order original packaging materials at 

your costs.

Mark the package completely

Enclose the completed health and safety clearance form.

Notify the carrier of any possible contamination if required.

Scrapping and waste disposal

Dispose of the equipment and any components removed from it safely in accordance 

with all local and national safety and environmental requirements. Particular care 

must be taken with components and waste oil which have been contaminated with 

dangerous substances from your processes. Do not incinerate fluoroelastomer seals 

and O-rings. You may authorize us to dispose of the equipment 

at

 

your expense. 

Otherwise we return the device at your expense.

Repair - Maintenance - Return - Calibration

IMPORTANT

Summary of Contents for BVC control

Page 1: ...nged It is strictly the users responsibility to check carefully the validity of this document with respect to his product 11 08 2015 999281 Technology for Vacuum Systems Decontamination of BVC basic B...

Page 2: ...d distributed completely and unchanged It is strictly the users responsibility to check carefully the validity of this document with respect to his product 11 08 2015 999281 After sales service Contac...

Page 3: ...ty to check carefully the validity of this document with respect to his product 11 08 2015 999281 Contents Safety information 5 Materials 6 Decontamination of the vacuum pump 8 Disassembling the BVC c...

Page 4: ...y the users responsibility to check carefully the validity of this document with respect to his product 11 08 2015 999281 WARNING indicates a hazardous situation which if not avoided could result in d...

Page 5: ...ng system Ensure that the device is decontaminated before repair and that you take adequate precautions to protect people from the effects of hazardous substances if contamination has occurred Take ad...

Page 6: ...FE PP PPS POM ETFE aluminium alloy Feet NBR BVC control G BVC professional G Support cover pump PBT Pump housing base plate aluminium alloy On off switch PC EPTR IEC plug PET Feet PU Screws stainless...

Page 7: ...ETFE Outlet ETFE Silencer silicone rubber Sensor PPS fuorosilicone Sensor tube PTFE PVC Collection bottle 2l Bottle borosilicate glass Cap insert PP Flat seal EPDM Hose nozzle at bottle head connecti...

Page 8: ...minated parts Adopt suitable measures to prevent the release of dangerous toxic explosive cor rosive noxious or polluting fluids vapors and gases and a possible contamination of personal and environme...

Page 9: ...aphragm key SW 46 order no 637777 Screw driver Torx TX20 Open ended wrench SW 19 Tweezers Connect person to ground before coming in contact with electronic parts IMPORTANT Disassembling the BVC contro...

Page 10: ...tly the users responsibility to check carefully the validity of this document with respect to his product 11 08 2015 999281 Unscrew hose nozzle or hose nozzle with coupling using an open ended wrench...

Page 11: ...ty of this document with respect to his product 11 08 2015 999281 View of the disassembled pump head parts ME 1C 1 Cover of housing cover 2 Housing cover 3 Housing cover insert 4 Valves 5 Head cover 6...

Page 12: ...m to its lowest oscillating position Unscrew di aphragm support disc with diaphragm and diaphragm clamping disc If the diaphragm is difficult to separate from the diaphragm support disc immerse assemb...

Page 13: ...ment with respect to his product 11 08 2015 999281 Lay down the BVC as shown in the figure Unscrew screws at the base plate Slide the base plate to the side Remove earth cables Remove hose from pressu...

Page 14: ...20 Documents are only to be used and distributed completely and unchanged It is strictly the users responsibility to check carefully the validity of this document with respect to his product 11 08 20...

Page 15: ...mpletely and unchanged It is strictly the users responsibility to check carefully the validity of this document with respect to his product 11 08 2015 999281 Disassembling the BVC basic Disassemble th...

Page 16: ...users responsibility to check carefully the validity of this document with respect to his product 11 08 2015 999281 Disassembling the VHCpro Turn cap 1 and remove 2 Pull adapter with hose out of the...

Page 17: ...pletely and unchanged It is strictly the users responsibility to check carefully the validity of this document with respect to his product 11 08 2015 999281 It is not possible to disassemble the VHCpr...

Page 18: ...no original components Drain the device completely of fluids and residues Decontaminate the device Close all openings airtight especially if using substances hazardous to health To expedite repair an...

Page 19: ...r service staff 11 If the paint is damaged we wish a repaint or a replacement of parts for reason of appearance repaint and replacement at customer s expense yes no 12 Legally binding declaration We a...

Page 20: ...btained under lab conditions to your specific operations depends on a number of factors beyond our control It is therefore strictly the users responsibility to very carefully check the application of...

Reviews: