3. Specification
The meter shall be designated for one year and recalibrated at 18 ° C to 28 ° C and a relative
humidity of less than 75%.
3.1 Summary
Automatic range and Manual range
Full ranges overload protection.
The maximum voltage allowed between the measuring end and the ground: 600V DC or
AC 600V RMS
Fuse protection: FF10A/250V on A
Working height: max 2000m
Maximum display value: 6000 counts
Polarity indication: automatic indication, '-' indicates negative polarity.
Overrange display: '0L' or '-0L'.
Sampling time: about 3 times / sec
Unit display: with function, power unit display.
Automatic shutdown time: no letter when about 15 minutes
Power supply: DC power supply 4.5V
Battery Type: AAA1.5Vx3 Battery
Battery under voltage indication: LCD display
-
symbol .
Temperature coefficient: less than 0.1 × accuracy /
℃
Operating temperature: 18
℃
~ 28
℃
Storage temperature: -10
℃
~ 50
℃
Size: 140
(
L)×
68
(W)×4
9(H)mm
Weight: About 226g (including battery)
3.2
Technical indicators
Ambient temperature: 23 ± 5
℃
Relative humidity: <75%
3.2.1
True RMS zero input
3.2.1.1 The measurement of sine wave signal, the use of true RMS measurement method than