![UWT NivoRadar NR 7100 Series Technical Information/Instruction Manual Download Page 13](http://html1.mh-extra.com/html/uwt/nivoradar-nr-7100-series/nivoradar-nr-7100-series_technical-information-instruction-manual_1066137013.webp)
NivoRadar
®
Continuous level measurement
Series NR 7100
Technical information / Instruction manual
Technical data
page 13
Ʋ
Max. spectral transmission power
density
+34 dBm/50 MHz EIRP
Ʋ
Max. power density at a distance of
1 m
< 3 µW/cm²
Ambient conditions
Ambient temperature
-40 … +60 °C (-40 … +140 °F)
Storage and transport temperature
-40 … +80 °C (-40 … +176 °F)
Mechanical environmental conditions
Vibrations (oscillations)
Class 4M8 acc. to IEC 60271-3-4 (5 g at 4 … 200 Hz)
Impacts (mechanical shock)
Class 6M4 acc. to IEC 60271-3-6 (50 g, 2.3 ms)
Impact resistance
IK07 acc. to IEC 62262
Process conditions
For the process conditions, please also note the specifications on the type label. The lowest
value (amount) always applies.
Process temperature
-40 … +60 °C (-40 … +140 °F)
Process pressure
-1 … 3 bar (-100 … 200 kPa/-14.5 … 43.51 psig)
Electromechanical data
Cable entry
Ʋ
Options
M20 x 1.5; ½ NPT
Ʋ
Cable gland
M20 x 1.5 (cable diameter 5 … 9 mm)
Ʋ
Closing cap
½ NPT
Wire cross-section (spring-loaded terminals)
Ʋ
Massive wire, stranded wire
0.2 mm² (AWG 24) … 2.5 mm² (AWG 14)
Ʋ
Stranded wire with end sleeve
0.2 mm² (AWG 24) … 1.5 mm² (AWG 16)
Bluetooth interface
Bluetooth standard
Bluetooth 5.0
Frequency
2.402 … 2.480 GHz
Max. emitted power
+2.2 dBm
Max. number of participants
1
Effective range typ.
1)
25 m (82 ft)
Adjustment
Smartphone/Tablet
Adjustment app
1) Depending on the local conditions