MultiDoc-It Imaging System
9
81-0214-01 Rev L
Service Procedures
Return Procedure
A
Returned Goods Authorization (RGA)
number must be obtained from UVP Customer Service
before returning any product.
Replacement Parts and Accessories
Accessories are available for use with the MultiDoc-
It Imaging Systems. Contact UVP’s offices at the
telephone numbers below for replacement parts for MultiDoc-It equipment. For transilluminator
replacement parts, refer to the separate transilluminator manual.
Description
Part Numbers
UV Lamps:
UVG-11, 254nm UV, 115V
95-0016-14
UVG-11, 254nm UV, 230V
95-0016-15
UVGL-15, 365/254nm, 115V
95-0017-09
UVGL-15, 365/254nm, 230V
95-0017-10
UVL-21, 365nm, 115V
95-0018-02
UVL-21, 365nm, 230V
95-0018-03
UVGL-25, 365/254nm, 115V
95-0021-12
UVGL-25, 365/254nm, 230V
95-0021-10
UV to 480nm Filter Converter
76-0310-01
Doc-ItLS Analysis Software
97-0185-02
Thermal Printer, Mitsubishi Digital (115/230V)
89-0069-06
Thermal Printer, Mitsubishi Digital (230V)
89-0069-07
Thermal Printer, Sony Digital (230V for Europe)
98-0357-01
Thermal Paper, matte 4 rolls (800 images)
89-0038-01
Thermal paper, 16 rolls (3200 images)
89-0038-04
Thermal Paper, 40 rolls (8000 images)
89-0038-05
UV
to White Light Converter Plate (21x26cm)
38-0191-01
UV to White Light Converter Plate (20x40cm)
38-0191-02
Visi-Blue Converter Plate (21x26cm)
38-0200-01
Visi-Blue Converter Plate (20x40cm)
38-0200-02
Filter, 50mm, sq., SYBR Green and EGFP
38-0219-01
Filter, 50mm, sq., Ethidium Bromide
38-0220-01
Fuse, 2 Amp, 250 V (Qty. 2)
56-0002-01
Fuse Cap (Qty. 2)
56-0036-01
Troubleshooting
No Power to the Darkroom
1. Recheck main power cord connections to the darkroom.
2. Check fuses, located at the back of the unit, near the power port. A flat-head screwdriver
will be required. Turn the fuseholder cap counterclockwise and the fuse holder will pop
out. Inspect the thin wire within the glass fuse to see if there is a break in the wire. If so,
replace the fuse(s). If fuses are blowing repeatedly, contact UVP Technical Support for
additional troubleshooting.
Transilluminator Will Not Turn On
1. There is a UV interlock switch that turns off the transilluminator and epi UV illumination
when the darkroom door is opened. Be sure the darkroom door is completely closed.
2.
Be sure the darkroom cabinet’s “Power On” indicator light is lit. If not, refer to “No Power
to the Darkroom” above.