MultiDoc-It Imaging System

 

 

81-0214-01 Rev L 

System Operation   

Darkroom Operation 

1.  Turn on the green darkroom main power switch, located at the top of the MultiDoc-It darkroom. 

The main power switch controls the automatic transilluminator shut off switch and the white 
light power. 

2.  To use the transilluminator, once the main power switch is on, turn on the green 

transilluminator power switch. If the darkroom door is opened, the transilluminator 
automatically shuts off to prevent accidental UV exposure.  

3.  To use the overhead white light, switch the 

Overhead Light

 button to 

White. 

Loading Samples Into the Darkroom 

Samples are loaded into the MultiDoc-It darkroom through the front access door.   

1.  Turn 

Off 

the UV Transilluminator. 

2.  Open the darkroom door. 

 

3.  Place the sample to be imaged in the center of the transilluminator filter glass. 

 

4.  Close the darkroom door. 

 

5.  Turn 

On

 the appropriate light source (UV transillumination, epi overhead white light or optional 

epi overhead lamp modules).   

Note:

 It is important to minimize the light source exposure to the sample. Turn off the light source as 

soon as the image is captured.   

View a Sample Prior to Capture  

Samples may be viewed through the Gel Viewer Window when the main darkroom door is closed. 

1.  Open the Gel Viewer Window on the darkroom door. 

2.  After visually verifying that the sample is loaded correctly, close the Viewer. Otherwise, light 

from the Viewer Window will interfere with the image capture process. 

Move or Cut a Sample 

It is possible to cut or move the sample by using the side access doors on the darkroom (if equipped). 

1.  Open the darkroom door and place the sample on the transilluminator.  

2.  Close the door and turn the transilluminator on. 

3.  Open the Gel Viewer Window on the darkroom door. 

4.  Open the side access door(s) and move or cut the gel while viewing the sample through the 

Viewer window. 

Software Operation 

1.  Open the Doc-ItLS software. 

2.  A window displaying the camera options should appear as pictured below. If not, click on the 

Acquisition Action tab and then on the Camera menu button. 

 

Summary of Contents for MultiDoc-It

Page 1: ...er Instructions UVP LLC Ultra Violet Products Ltd 2066 W 11th Street Unit 1 Trinity Hall Farm Estate Upland CA 91786 Nuffield Road Cambridge CB4 1TG UK Phone 800 452 6788 909 946 3197 Phone 44 0 1223 420022 Fax 909 946 3597 Fax 44 0 1223 420561 Web Site www uvp com ...

Page 2: ...It models include side access doors for moving or cutting gels without opening the front door UV Filter The included emission filter an orange colored UV blocking bandpass filter is used to absorb UV and IR radiation from the transilluminator and to enhance the orange pink bands generated by Ethidium Bromide stained gels The filter can be removed when imaging non fluorescent media protein gels col...

Page 3: ... Vista or XP SP2 or later Internet Explorer 6 0 or higher To locate the version of Internet Explorer open Internet Explorer and click on Help About Internet Explorer Intel Pentium Processor or equivalent 1 6 GHz or higher 1 GB of RAM or greater 200 MB of available hard disk space for the program more for data CD ROM drive One Universal Serial Bus USB for connecting the camera Color monitor support...

Page 4: ... back of the darkroom 2 Attach the main power cord to the top rear of the darkroom and the other end of the main power cord into a power outlet a surge suppressor is recommended Figure 1 Filter Installation 1 Unscrew the four plastic head screws Fig 1 that hold the filter tray to the camera mounting plate 2 Remove the filter tray from its position and place the filter s in the square recessed area...

Page 5: ...osing Restart from the Windows Start menu 5 Double click onto the UVP icon on the desktop 6 Activate the software by choosing On the fly activation if connected to the Internet Complete all the required information on the form and fill in the Serial Number located on the back of the flash drive s tin box Click onto Get Activation No and then click onto Activate when the Activation Number appears i...

Page 6: ...i overhead white light or optional epi overhead lamp modules Note It is important to minimize the light source exposure to the sample Turn off the light source as soon as the image is captured View a Sample Prior to Capture Samples may be viewed through the Gel Viewer Window when the main darkroom door is closed 1 Open the Gel Viewer Window on the darkroom door 2 After visually verifying that the ...

Page 7: ...us radio button on the Preview window then use the left and right arrows shown in the red box below to adjust the focus settings Preview Window with Manual Focus 7 When satisfied with the previewed image select Capture at the top of the software Camera Setup Image Acquisition If operating the camera in automatic mode is not sufficient for the application try utilizing the manual mode to gain addit...

Page 8: ...sed Separately 1 Move the UV Lamp power switch to the ON position 2 Slide the UV Lamps into the lamp brackets inside the darkroom Be sure the lamp power cords are positioned toward the outlets located on back inside of the darkroom 3 Run the lamp power cords over the camera mounting plate and plug into the outlets 4 To use the lamps the darkroom main power switch must to be in the ON position and ...

Page 9: ... images 89 0038 01 Thermal paper 16 rolls 3200 images 89 0038 04 Thermal Paper 40 rolls 8000 images 89 0038 05 UV to White Light Converter Plate 21x26cm 38 0191 01 UV to White Light Converter Plate 20x40cm 38 0191 02 Visi Blue Converter Plate 21x26cm 38 0200 01 Visi Blue Converter Plate 20x40cm 38 0200 02 Filter 50mm sq SYBR Green and EGFP 38 0219 01 Filter 50mm sq Ethidium Bromide 38 0220 01 Fuse...

Page 10: ...e for one 1 year from date of purchase Consumable and disposable parts including but not limited to tubes and filters are guaranteed to be free from defects in manufacture and materials for ninety 90 days from date of purchase Transilluminators carry a two 2 year warranty from date of purchase If equipment failure or malfunction occurs during the warranty period UVP shall examine the inoperative e...

Reviews: