Uvitron International PORTA-RAY 400/600W Instruction Manual, Rev. E
Page 4 of 22
INSTALLATION
2
2.1 General
The PortaRay light cure system may be placed on any bench or shelf that has access to an
120VAC or 240VAC 50/60-cycle single-phase power source in a clean operation area.
CAUTION:
Unless otherwise specified at time of order, the system ships with the
default configuration of 100/120VAC ±10%. To order systems configured for
200/240VAC ±10%, contact the factory or your
Uvitron representative (refer to Configurations table
in Specifications section for part numbers). The
unit
’s voltage configuration is clearly identified by a
bright orange label and also on the serial number
label, both at the side of the lamp head. Operating
with voltage outside the specified range may result
in damage to the system.
Since the PortaRay is an air-cooled unit, dust or airborne
particles can clog the internal cooling passages and cause
overheating. Allow a 6-inch clearance on all sides of the
lamp head for unrestricted clean airflow.
2.2 Mounting
The lamp head assembly should always be placed on the headrest, or held away from any
flammable surfaces. Never place the lamp head directly on the work surface. The PortaRay is
also designed to be fixture, machine or conveyor mounted using the threaded holes in the top
or bottom surface of the lamp head (see fig. 2 below).
Fig. 2, PortaRay lamp head top and bottom side mounting hole dimensions (in inches).
Fig. 1, PortaRay lamp mounted
on closed type headrest