UVCpectra Standalone ULTRA 300 User Manual Download Page 9

 

Page | 9 
 

9- Maintanence 

 

UVC Fluorescent light bulbs have 9.000h Useful Life Time. 
 
Please call technical service to change the UVC Bulbs and Reflection system when 
9.000 hours is reached. 
 

LIGHT TECHNICAL 

 

 

Main Application  

Disinfection 

Useful Life (Nom)  

9000 h 

System Description  

High Output 

Color Code  

TUV 

Depreciation at Useful Lifetime  

15% 

UV-C Radiation at 100 hr  

25 W / bulb 

 

 

10- Possible Faults and Troubleshooting 

 
Consult an authorised technician if RED Blinking 
 
You can see the error from the user interface. 
 

LIGHTS 

Meaning 

Action 

RED LIGHT BLINK 

FAN or Bulb error 

Goto user interface and see error 

 

 

 

 

 

If there is a fan error start the unit with 100% fan speed 

 

 

If fan error continues, call technical center 

 

 

 

 

 

If there is a bulb error call technical center for replacement 

 

 

 

 

 

 

GREEN LIGHT BLINK 

Cannot access Wifi 

TURN Off the system with ON /OFF switch 

 

 

Wait 10 seconds 

 

 

TURN ON the system 

 

 

 

 

 

If problem continues, Restart your WiFi access point and try again 

 

 

 

 

 

If problem continues, RESET WiFi settings to factory defaults. 

 

 

Press and HOLD WiFi button 10 seconds. 

 

 

Do the installation again 

 
 
 

 

 

Summary of Contents for Standalone ULTRA 300

Page 1: ...Page 1 UVCpectra TUBE ULTRA 300 UVCpecta StandAlone ULTRA 300 USERS MANUAL Version 1 1 All rights reserverd Users manual cannot be changed or copied...

Page 2: ...Page 2...

Page 3: ...tive Signs 4 4 Standards 4 5 Safety Instructions 5 6 Handling and Carrying Information 5 7 Installation and assembly 6 8 Usage Instructions 6 9 Maintanence 8 10 Possible Faults and Troubleshooting 8 1...

Page 4: ...0 UVCpectra Stand Alone Ultra 300 Dimensions 160 x 255 x 1470mm 600 x 400 x 1600mm Electrical Power 230 Watts 230 Watts Working Voltage 220 230 V 220 230 V Frequency 50 Hz 50 Hz IP Rating IP 54 IP 54...

Page 5: ...achine before reading users manual WARNING Danger Sign Information Sign Protective Ground Connection High Voltage Hazard Electrics Electronics Unit Do not throw to waste bin UV IRRADIATION DANGER DO N...

Page 6: ...remove the parts preferably with gloves Put them in a sealed plastic bag and take it to your local waste facilities for recycling Do not use a vacuum cleaner DANGER Risk Group 3 Ultra Violet product...

Page 7: ...XX 5 Default WiFi key is uvcpectra 6 If the IP address is not automatically assigned to 192 168 4 X manually change your computers IP address to 192 168 4 10 7 Use a Chrome browser to log in to http 1...

Page 8: ...r Interface Optional If you want to access the User Interface from Internet consult your IP admin to make a port forwarding Example Forward Port 1101 to my UVCpectra Now you can access with a browser...

Page 9: ...echnician if RED Blinking You can see the error from the user interface LIGHTS Meaning Action RED LIGHT BLINK FAN or Bulb error Goto user interface and see error If there is a fan error start the unit...

Page 10: ...not in warranty Add1 Spare Parts List Tube 300 Ultra and Standalone 300 Ultra UVC Bulb UVCpFLph75300 Driver 1 1x80Watt UVCpDRos80x1 Driver 2 2x80 Watts UVCpDRos80x2 Fan UVCpFNruEC1502 EC Board UVCpECo...

Page 11: ...Page 11 Add2 EU Conformity Declaration...

Page 12: ...Page 12 Add3 Electrical Schema Button LED Configurations...

Reviews: