Page | 9
9- Maintanence
UVC Fluorescent light bulbs have 9.000h Useful Life Time.
Please call technical service to change the UVC Bulbs and Reflection system when
9.000 hours is reached.
LIGHT TECHNICAL
Main Application
Disinfection
Useful Life (Nom)
9000 h
System Description
High Output
Color Code
TUV
Depreciation at Useful Lifetime
15%
UV-C Radiation at 100 hr
25 W / bulb
10- Possible Faults and Troubleshooting
Consult an authorised technician if RED Blinking
You can see the error from the user interface.
LIGHTS
Meaning
Action
RED LIGHT BLINK
FAN or Bulb error
Goto user interface and see error
If there is a fan error start the unit with 100% fan speed
If fan error continues, call technical center
If there is a bulb error call technical center for replacement
GREEN LIGHT BLINK
Cannot access Wifi
TURN Off the system with ON /OFF switch
Wait 10 seconds
TURN ON the system
If problem continues, Restart your WiFi access point and try again
If problem continues, RESET WiFi settings to factory defaults.
Press and HOLD WiFi button 10 seconds.
Do the installation again