Upstream Instruction Manual
Released on 5/1/10 – R800001L
Page 32
UV Pure Technologies Inc. © 2010
Tools Needed
Slotted screwdriver
3/16” hex wrench (Allen wrench)
NOTE:
Resetting the Lamp Lifetime counter will clear the Message History,
Lamp Starts counter and Power Ups counter. If this information is required,
review it first before proceeding to Step 1.
www.uvpure.com
V
U
FACES OUT
FACES IN
INCORRECT TAB ORIENTATION
(TABS FACING INWARDS)
CORRECT TAB ORIENTATION
(TABS FACING OUTWARDS)
USE A SCREW
DRIVER TO
PRY LAMP
UP HERE
NOTE: DO NOT TWIST/TURN THE LAMPS
WHILE PINS ARE ENGAGED
UNPLUG
UNIT
TO OPEN
LAMP
SOCKETS
TOP COVER
WING NUT
OR
THUMB
SCREW
UV LAMP
FRONT PANEL
ASSEMBLY
Figure 3
Step 1:
From the Main Menu, move down to
Lamp life remaining
screen. Press
Enter to move to next screen to reset counter. Press Enter again and then push
Up or Down key to select YES and then press Enter. The lamp hours will reset to