5 Specifications
Non-invasive upper-arm blood pressure
•
Oscillometric method using step controlled
deflation
•
Adult and pediatric patients
•
BP Measurement range:
•
Systolic: 40 to 250 mmHg
•
Diastolic: 20 to 190 mmHg
•
Mean: 25 to 245 mmHg
•
Cuff Pressure: 0 to 300 mmHg
•
BP Measurement accuracy:
Static
Accuracy: ± 3 mmHg
•
Clinical Accuracy: Mean error is less
than ± 5 mmHg Standard Deviation is
less than 8 mmHg
•
Pulse Rate range: 40 to 200 bpm
•
Pulse Rate accuracy: greater of
±
3% or
±
3bpm
•
Power-on self-test
•
Typical measurement time within 45
seconds
Non-invasive central blood pressure
•
Model-based transform of peripheral
waveform reports Systolic and Diastolic
central pressure
Pulse wave measurement
•
Arterial measures:
•
Peripheral Augmentation Index (sAI)
•
Suprasystolic Reflected Wave Transit
Times, peak and foot (sRWTT
p,
and
sRWTT
f
)
•
Cardiac function measures:
•
Maximum Suprasystolic pressure
gradient (sdP/dt)
•
Suprasystolic Pulse Pressure Variation
(sPPV)
•
Systolic Ejection Period (sSEP)
•
Suprasystolic Pulse Rate Variability
(sPRV)
•
Signal to Noise ratio (SNR)
•
Data Sampling Rate: 200Hz
Physical dimensions
•
Size: 156 mm × 157 mm × 119 mm
•
Weight: 500 g
Display and memory
•
320×240 pixel matrix 3.5” TFT colour
display
•
Pulse wave waveform display
•
Graphical rhythm strip display
•
Status and error display
•
SD Card slot, hot swappable with cards up
to 32 GB
Blood pressure cuff and hose
•
Lifetime antimicrobial treatment
•
Small Adult, Adult and Large Adult cuffs
available
•
1.3 m hose
Computer interface
•
Computer interface specification (serial
port) available for customer configuration
Power supply
•
Power Supply Unit: GlobTek Inc model
GTM41060-2512 medical grade, AC/DC
wall adapter
•
AC Input: 100–240 V
AC
, 50–60 Hz, 0.6 A
•
DC Output: 12 V
DC
, 2.08 A
•
Use of the BP+ with the Power Supply
Unit is considered an ME system
Environmental specifications
•
Operating Temperature: 10 to 40 °C (50
to 104 °F)
•
Operating Relative Humidity: 15 to 80%
(non-condensing)
•
Operating Atmospheric Pressure: 80 to
103 kPa
•
Transport and Storage Temperature: -20
to 50 °C (-4 to 122 °F)
•
Transport and Storage Relative Humidity:
15 to 80% (non-condensing)
IEC 60601-1 standard classification
•
Class II Equipment
•
Type BF Applied Part (the blood pressure
cuff is considered the applied part)
•
Continuously powered with
measurements being taken upon demand
•
Equipment not suitable for use in the
presence of a flammable anesthetic
mixture, with air or oxygen or nitrous
oxide
Warranty
•
12 month warranty on the device
•
6 month warranty on cuffs and extension
hose
Reference to standards
•
EN 1060-1 /-3 /-4; IEC 60601-1;
IEC 60601-1-2 (EMC); IEC 60601-1-11