4.6 Error Message
4.6.1 Main Screen Message
Message
Reason
User Action
Connection Failed
or
No Connection
Internal fault condition.
Consult Uscom / Uscom
certified servicing technician
Wrong power adapter used.
Check the power adapter
being used is the one
provided with the device, as
specified in Section
5
Specification
Measurement Timeout
Measurement could not be
completed within the maximum
permitted time. Safety system has
cancelled the measurement.
Redo another measurement.
Ask the patient to relax, sit
still, not to talk, no arm or
finger movement on the arm
being measured (left), etc.
Refer to section
2.1
&
2.2
.
If error persists, consult
Uscom / certified servicing
technician (Please make a
note of any error number
displayed along with the error
message.)
Systolic BP too High
or
Unable to measure BP:
Range Exceeded (####)
(Where #### is the error
number)
The upper arm systolic pressure
measured is too high. The device
cannot proceed to do the Pulse
Wave measurement
Unable to Calculate
Results
Unable to Find Features
Unable to Find Pulse
Wave
Calculation error
Unable to Reinflate
The inflation target for doing the
Pulse Wave measurement is too
high
Unable to Start
Device detects an internal error
when it starts the blood pressure
measurement
Invalid Signal
The signal noise ratio is too low and
out of the valid range.
Poor Signal
The signal quality is poor
Unable to measure BP:
Please Repeat (####)
(Where #### is the error
number)
Device detects a blood pressure
measurement related error when it
is doing the blood pressure
measurement
Unable to measure BP:
Timeout (####)
(Where #### is the error
number)
Blood pressure measurement
Timeout
Unable to measure BP:
(####)
(Where #### is the error
number)
Blood pressure measurement failed Consult Uscom/Uscom
certified servicing technician
with the BP error number
Page 26 of 31