7.0 Technical
Data
(at 23° C ± 5° C, max. 75 % rel. humidity). Temperature coefficient (<18°C or >28°C):
0.15 (Specified Accuracy) /°C
Display:
4digit, LCD incl. functions and symbols
Range Selection:
automatic
Auto-Power-Off:
approx. 30 min.
Overload display:
OL
Measuring rate:
2 measurements/sec. (LCD)
Clamp opening:
ca. 30 mm (1.2 inch)
Overvoltage Categorie:
CAT II, 600 V / CAT III, 300 V
Pollution degree:
2
Height above MSL.:
up to 2000 m
Battery display:
at low battery
Power supply:
Batterie 2 x 1,5 V IEC LR6 (alkaline)
Current Consumption:
approx. 10 mA
Operation temperature:
-10° C … 50° C at less than 85% rel. H.
Storage temperature:
-20° C … 60° C at less than 75% rel. H.
Dimension:
210 x 62 x 35,6 mm
Weight:
approx. 200 g
Voltage AC TRMS
Crest Faktor <4, Input Impedance 10M
Ω
, max. Peak Voltage 850 Vss
Range
Resolution
Accuracy
5 V...600 V
0,1 V
± 0,5 % ± 5 D (50/60Hz)
20 V...600 V
0,1 V
± 1,5 % ± 5 D (45-1kHz)
Overvoltageprotection: 800V AC
Current AC TRMS (mA and A)
Crest Faktor <4, max. Peak Voltage 850 mAss (mA Range) or
70 Ass (A Range)
Range Resolution
Accuracy
(50/60Hz)
0,50...60,00 mA
0,01 mA
± 0,5 % rdg. +5 D
60,00...99,99 mA
0,01 mA
± 0,5 % rdg. +50 D
100...600,0 mA
0,1 mA
± 0,5 % rdg. +5 D
0,050...3,000 A
0,001 A
± 0,5 % rdg. +5 D
3,000...9,99 A
0,001 A
± 0,5 % rdg. +50 D
10,00...30,00 A
0,01 A
± 0,5 % rdg. +5 D
30,00...50,00 A 0
0,01 A
± 1,0 % rdg. +5 D
14
PTDB93570002.qxd:Layout 1 30.05.2008 15:05 Uhr Seite