DNx-AI-208 Analog Input Layer
21
Tel: 508-921-4600
www.ueidaq.com
Vers:
4.6
Date: November 2013
File:
AI-208 App A.fm
© Copyright 2009
United Electronic Industries, Inc.
shows a typical single-channel wiring diagram for a Quarter-bridge
Strain Gauge connected to the STP-AI-208 panel. As the figure indicates, you
should remove the board-mounted jumper when you use a 3-wire Circuit. Note
that a quarter-bridge circuit requires that you solder precision resistors to the
board where indicated in
to complete the measuring bridge. As an
alternative, you can install precision resistor-divider networks in SOT23 pack-
ages directly on the board, as shown in
, to complete the bridge cir-
cuits.
Figure A-5 Single-Channel Wiring Diagram — Quarter-Bridge
!
#
"$#$%"!&$%
'$
Insert jumper when using a
2-wire connection. Remove
when using a 3-wire connection.
User-supplied
bridge-completion
resistors (Soldered
onto board or connected
to screw terminals.)
User-supplied
bridge-completion
resistors (resistor-
divider in SOT23
package, soldered
to board.)