14
MEASUREMENTS
Hydrogen sensors can be used for a wide variety of
measurements (see our web page for further information www .
unisense .com) . The most common use of hydrogen sensors is for
making profiles in e .g . sediment or animal tissue where a high
spatial resolution is wanted, or for hydrogen measurements in
water samples .
m
ounting
of
tHe
sensors
Although the Unisense microsensors are made of glass, the
tip is flexible and can bend slightly around physical obstacles .
The sensor is thus rather sturdy in the longitudinal direction .
However, large obstacles like stones or lateral movements of the
sensor when the tip is in contact with a solid substrate may cause
the tip to break .
Furthermore, due to the small size of the microsensor tip and
to the steepness of gradients in many environments, even a
displacement of the sensor tip of few microns may change its
environment .
Therefore, we recommend that measurements should be
performed only in a stabilized set-up free of moving or vibrating
devices . We recommend the Unisense lab stand LS and the
Unisense micromanipulator MM33 (MM33-2 or MMS) for
laboratory use . For in-situ use, we recommend our in situ stand
(IS19) and a micromanipulator .
e
lectrical
noise
The signal of the microsensor is very small (10
-13
to 10
-10
ampere) .
Although both the Unisense picoammeters and the Unisense
Hydrogen microsensors are very resistant to electrical noise from
the environment, electrical fields may interfere with the sensor
signal . Therefore we recommend that unnecessary electrical/
mechanical equipment is switched off and the sensor or wires are
not touched during measurements and signal recording .
Micromanipulator
Summary of Contents for HYDROGEN SENSOR
Page 1: ...1 HYDROGEN SENSOR USER MANUAL ...
Page 2: ...Hydrogen sensor user manual Copyright 2016 Unisense A S Version December 2016 ...
Page 3: ...HYDROGEN SENSOR USER MANUAL UNISENSE A S ...
Page 4: ......
Page 22: ...22 ...
Page 23: ......
Page 24: ...UNISENSE DENMARK www unisense com info unisense com ...