18
4
Torque Settings
18
Torque Settings
Chapter
4
4-3.
Calibration procedure for rotation speed
Set the type of UTM to be used.
Set the unit of value to display.
Set the minimum unit of rotation speed such as scale
interval and scale division.
(optional if there is no change)
Choose whether or not to use the function to forcibly
reset the rotation speed to 0 in an quick stop.
(optional if there is no change)
UTM type (UTM TYPE)
Set the type of UTM to be used.
<Setting range>
UTM,
UTM
Ⅱ
Unit (UNIT)
Set the unit of value to display.
<Setting range>
rpm, min
-1
Minimum scale (MIN. SCALE) (optional if there is no change)
Set the minimum unit of rotation speed such as scale interval and scale division.
<Setting range>
1, 2, 5, 10, 20, 50, 100
Quick stop mode (QUICK STOP) (optional if there is no change)
This is a function to shorten the time for the rotation speed display of the TM400 to reach zero after
rotation stop.
<Setting range>
OFF, ON
- OFF
The display reaches zero about 1s after rotation stop.
- ON
The display is forcibly reset to zero in case of rapid deceleration (The rotation speed
is reduced rapidly to 1/4.).
*
:
Be sure to perform the setting.
(1)
UTM type
*
(2)
Unit
*
(4)
Quick stop mode
(3)
Minimum scale