34
5
FUNCTION SETTINGS
34
FU
N
C
TIO
N
S
ET
TIN
G
S
Chapter
5
■
Setting of motion detection parameters
- Motion detection mode (setting mode 1-5)
elect the stable condition from stable mode and checker mode.
- Motion detection period (setting mode 2-3)
Set the period to judge that the weight value is stable.
(Input range/0.0 - 9.9)
- Motion detection range (setting mode 2-3)
The variation width of the weight value is compared with this setting value multiplied by the min
scale division.
(Input range/0 - 99)
5-7.
Zero tracking(ZT) (setting mode 2-4, 2-5)
This function automatically corrects slow zero drifts and minute zero point displacements caused
by weighing residue, mud, dust, snow, and other accumulations.
- Zero tracking period (setting mode 2-4)
Input range/0.0 - 9.9
- Zero tracking range (setting mode 2-5)
Input range/0 - 9999
Weighing function 2
Stable detection mode
0: Stable mode
1: Checker mode
Range (00 - 99)
Period (0.0 - 9.9)
Motion detect
(period - range)
- Zero tracking is the function to automatically zero the gross weight when the
state in which zero point displacements are within the set tracking range
continues for the set period or more.
- Set the zero tracking period in the range of 0.0 - 9.9 sec, and zero tracking range
(digit) in the range of 0 - 9999 in units of 1/4 of weight display. (Setting values of
0002 and 0012 are equivalent to 0.5 and 3, respectively.) Also, when the period
is set at 0.0 sec and the range is set at 0000, zero tracking does not work.
Point
Boundary of zero track
Zero tracking period Indicated value
Zero tracking range
+
0
-
Zero tracking period
Resumed from the point of time of returning within the range