4.CALIBRATION
26
4-2. Actual Load Calibration
The actual load calibration uses the following procedure
Release the calibration LOCK that inhibits calibration.
Set the minimum value of the digital change desired.
(
omissible if no desired value is present
)
.
Set the zero point of the strain gauge sensor.
Set the span(gain) point of the strain gage sensor.
Set the calibration of ON to prevent misoperation.
Releasing Calibration LOCK
Setting of Scale Division
Zero Calibration
Actual Load Calibration
Calibration LOCK