3.DATA EXCHANGE WITH CPU
9
3-2-3. IN (F130
→
CPU Unit)
Digital zero response
Returns the state of the digital zero request bit.
Hold response
Returns the state of the hold request bit.
Hold value clear response
Returns the state of the hold value clear request bit.
A/D input range over
This bit becomes "1" when the input from the strain gauge type sensor is out of the input range of
the A/D converter.
Measurement value overflow
This bit becomes "1" when the indicated value is smaller than -32768, or larger than +32767.
Normally ON
Always returns "1."
Survival confirmation response
Returns the state of the survival confirmation request (n) 07 as it is inverted.
Zero calibration in progress
This bit becomes "1" while Zero calibration is in progress.
Actual load calibration in progress
This bit becomes "1" while Actual load calibration is in progress.
Equivalent input calibration response
Returns the state of the Equivalent input calibration request bit.
Calibration error occurrence
This bit becomes "1" if Zero calibration, Actual load calibration, or Equivalent input calibration
does not end properly with a calibration error.
Setting error occurrence
This bit becomes "1" when the initial set value assigned in the data memory area is out of the setting range.
Factory-adjusted operation response
Returns the state of the factory-adjusted operation request bit.