Please read the IFU carefully before fitting. Only correct usage will warrant the
function.

1. Intended Use
The knee joint JT40 is only suitable for prosthetic care of the lower extremities. The JT40
contains an effective hydraulic swing phase control. The back shifted pivot point ensures the
safety in the stance phase. The knee joint is intended for Mobility level 3 and 4 and is limited
to a body weight of max 100 kg.

Additional loads and patient activities must be taken into account when
calculating the load.

2. Technical Data

Proximal connection: pyramid
Distal connection:

30 mm Tube clamp

3. Indications/ Contraindications
Indications:

Transfemoral amputation

Knee disarticulation

Hip disarticulation

Congenital limb deficiency

Contraindications:

The patient is over 100 kg (220 lbs) and/or mobility level 1-2

The doctor decides the prescription of the product is inappropriate.

4. Side effects
There are no known side effects.

Order
No.

Material

Installation height

Weight

Flexion angle

Article No.

JT40

Aluminum

140 mm

496 g

170°

4 610 070 00 00 000

3

5. General Instructions

This medical device is designed for single patient, multiple use.

Fitting/service of this medical device is only allowed by a certificated orthopedic
professional.

The professional should instruct the correct use of the device to the user.

The joint must be protected against dust, moisture, aggressive media, and mechanical
forces so its function is not affected.

The joint is intended for use at temperatures from -10°C to 50°C.

Do not longer use the knee joint if its function is impaired. This may be due to stiffness, 
restricted extension, noise generation, faulty swing phase control or stance phase safety, 
as well as impact from an external force (for example, if dropped). In such cases, it must 
be checked by a professional and, if necessary, sent to Uniprox GmbH & Co. KG.

There is a risk of jamming when the joint is in use.

The product must not be changed or applied in any improper way. Non-compliance can 
impair the function of the product and thus no liability will be accepted.

The safety of the user is not guaranteed if the product is not used for its intended purpose. 
This will also terminate the warranty of the product.

Joint components may heat up slightly during walking.

Silicone spray should be used to prevent noises being generated by contact between the 
cosmetic foam and knee joint. Do not use Talcum powder as it removes lubricant from the 
bearings and may lead to impaired function as a result.

Avoid loading onto the prosthetic knee joint in the maximum flexed position.

In case of maximum flexion, posterior part of the socket or other components may hit and 
damage the knee device and the socket (Fig. 1-a).  If it is inevitable, make a new hitting 
point on the distal part of the knee to reduce moment force (Fig. 1-b).

When the knee joint is bent, never put hands around the device.

Insertion of fingers between knee components or the knee device and another prosthetic 
component could cause severe injury such as laceration or fracture.

This instruction should be given to users as well.