ATMOS Operation Manual
page 18/24
Press the button “Open Valve”. You will hear a click.
Both pressure values will approach each other again. They should be very similar. If this is the case, the self
test of the device was successful and it runs correctly.
If the values differ, please run the test again, maybe with a different pipette.
If the test is still unsuccessful, please contact UniPix or your local sales person.