6. EXCHANGING BOARDS AND MODULES 

Service information Phyaction 190 / 190i - Page 17

 

 

 

CHAPTER 5

  EXCHANGING BOARDS AND MODULES

 

  

 
 

5.1 Introduction

 

The most frequent service operations will be the exchanging of boards or modules, because it would be too 
time consuming and costly to do component level service in the field. When exchanging boards or modules 
there are a few things that should be noticed. These attention points will be described in this chapter. 
 
If you want to exchange a module or a board you will have to open the appliance. It is probably easy to place 
the appliance on its rear side, with the bottom side towards you. You can unscrew the 5 screws of the bottom 
plate and tilt it towards you. If you want to test and make measurements you can leave the cables and 
connectors where they are, but if you want to exchange a board or module it is better to disconnect the cables 
and seperate the bottom plate from the rest of the appliance. 
 
If you have exchanged a board or module, always test the appliance. This functional test starts with switching 
the device on and checking the 4 LEDs, the fan and the LCD. If no error messages appear, check the memory 
(can you store a setting and retrieve it). Finally check the contact detection of the ultrasound heads. After this 
you should do a safety test (short description in section 3.2). 
 
In general it should be said that you should take the necessary precautions to prevent electrostatic discharges 
occur on the delicate electronic circuits of the unit. These precautions implicate that you should ground the unit, 
use a grounded desk overlay, ground the tools such as solder irons and you should be connected to ground by 
the usual wrist or ankle strap. 
 

5.2 

Exchanging the software or the main board (PCB 192X)

 

If you want to exchange the software or the main board of the appliance, you first open the appliance as 
described in section 5.1 

Introduction

. You disconnect the cable from the transformer to the main board and the 

cable of the fan. Then you remove the two screws from the main board, which are located at the back of the 
appliance (between the ultrasound module and the back side). You disconnect the cable from the main board to 
the keyboard and the cable of the backlight of the LCD. Now you can shift out the board and exchange the 
EPROM. If you want you can disconnect the other 4 cables, but that is not really necessary. If you want to 
exchange the main board you will have to disconnect the cables from console and from the LCD backlight, take 
out the board and disconnect the remaining cables. 
 
If you have exchanged the software or the main board, shift back the PCB on its place and connect all cables. 
Also connect the transformer cable. Leave the appliance on its rear side and switch it on. If any error message 
appears on the LCD, switch the appliance off and try to repair the cause of the error. After you have succeeded, 
you remove the external mains cable and fix the main board with the two screws. Then check whether all 
cables are connected (fan !!) and close the case. After you have screwed in the 5 screws on the bottom plate, 
you should do a short functional test and a safety test (short description in section 3.2). 
 

5.3 

Exchanging the ultrasound module

 

To remove the ultrasound module, you first remove the bottom plate as described in section 5.1 

Introduction

Make sure there are no treatment heads connected to the device. The easiest way to place the appliance is 
with the front side down. Put it on a soft underground to make sure you don't damage the front side. Disconnect 
the internal cable from the safety socket and unscrew the safety socket. Remove the screw to the tab at 
R82/C82, then unscrew the three other screws to release the ultrasound module. You can take the module out 
of the appliance case, remove the screws from the D-subminiature connector at the back of the module and 
disconnect the cable. 
 
If you install a new module there are a few things that you have to pay attention to. You start with connecting 
the cable to the module and fixing it with two screws. There are two O-rings between the front panel and the 
LEMO chassis parts (the sockets for the ultrasound output) to prevent any liquid from entering the appliance at 
that place. Make sure the O-rings are undamaged and are positioned in the chambers around the holes for the 
connectors. If you place the module you should make sure the LEDs are positioned in the LED-lenses. Before 
you fix the module with the screws you should lift the appliance to make sure the O-rings are still on their place. 

Summary of Contents for PHYACTION 190

Page 1: ...s a registered trademark of Uniphy BV Art Code 93001212 6 Phyaction is manufactured in The Netherlands by Uniphy BV P O Box 558 NL 5600 AN Eindhoven the Netherlands Tel 31 499 491800 Fax 31 499 474734...

Page 2: ...ion 9 3 2 IEC 601 1 safety tests 10 3 3 Performance check of the ultrasound treatment heads 10 CHAPTER 4 TROUBLE SHOOTING 12 4 1 Introduction 12 4 2 Description of the automatic selftest 12 4 3 Descri...

Page 3: ...sing the equipment in the vicinity of short wave or microwave equipment can influence the correct operation of the unit Using this equipment when high frequency surgical equipment is connected to the...

Page 4: ...ment and you must ensure that no liquid enters the equipment 1 3 5 Fitting the head holder As standard a head holder is supplied with each treatment head Before you use the Phyaction 190 for the first...

Page 5: ...n9 you can access the special memory function 20 different treatments can be saved retrieved or changed 1 4 5 Output connectors and sockets on the front The safety socket10 on the far left is used for...

Page 6: ...ppendix 2 TUNING THE ULTRASOUND TREATMENT HEADS the tuning process is described 1 4 9 Head holder For each Phyaction 190 a head holder for each treatment head is supplied as standard You can fit this...

Page 7: ...e complete apparatus The user can set most of the parameters with the keyboard which is shown at the left of the microcontroller The intensity of the ultrasound can be set with the intensity knob draw...

Page 8: ...connection to the UART of the microcontroller The interface is for extended service features used by manufacturer only 2 2 6 Power supply The power supply is a conventional one with a mains transforme...

Page 9: ...ulator in order to prevent erroneous frequencies being produced by the treatment head 2 3 2 Modulator To obtain a symmetrical steering signal for the power stage the modulator switches its outputs wit...

Page 10: ...is mainly a capacitive load Its impedance differs among other things with the size of the crystal and its operating frequency For this reason the microcontroller can match the impedance by switching o...

Page 11: ...cables and connectors undamaged Pay special attention to possible leaks in the treatment heads such as cracks or loosened seems Are the labels on the ultrasound treatment heads readable 3 1 2 Function...

Page 12: ...treatment heads In the software of the Phyaction 190 a special menu for service purposes only is included Most of these features can be used for trouble shooting You can find the description of the ot...

Page 13: ...cted The microcontroller will then measure the same parameter as before Vusi wet In the following table are the values for the different frequencies and treatment heads which should be met Head type f...

Page 14: ...described in the user manual Errors that cannot be corrected by the user are displayed as shown in figure 4 1 where the numbers are randomly chosen The error codes presented after Code are described...

Page 15: ...roller and its required hardware program and data memory 1 The ALU and a number of instructions are verified When an error is met ERROR 00 is generated and 6 beeps will sound 2 The internal memory the...

Page 16: ...gh it could indicate a hardware error in the head detection hardware Checking the ultrasound voltage 1 After power on of the unit the PWM is in its minimum position 0 255 Any voltage still present in...

Page 17: ...service personnel only so never tell a physiotherapist or doctor how to enter this menu because they might do things that will harm the appliance or the treatment heads If you press the yellow button...

Page 18: ...g ultrasound it will show a screen saver after 5 minutes of non activity After approximately 5 minutes the screen saver disappears and the LCD backlight is switched off To prevent the appliance to go...

Page 19: ...main board to the keyboard and the cable of the backlight of the LCD Now you can shift out the board and exchange the EPROM If you want you can disconnect the other 4 cables but that is not really nec...

Page 20: ...ew won t hold at all This would mean you would have to exchange the front panel which is at lot of work and a lot of money Finally connect the cable on the keyboard panel and re install the main board...

Page 21: ......

Page 22: ......

Page 23: ......

Page 24: ......

Page 25: ......

Page 26: ......

Page 27: ......

Page 28: ......

Page 29: ......

Page 30: ......

Page 31: ......

Page 32: ......

Page 33: ......

Page 34: ......

Page 35: ......

Page 36: ......

Page 37: ......

Page 38: ......

Page 39: ...ow key pressed until the Service Menu appears see figure 4 7 Now press function key 1 Set head parameters Figure 5 appears on the screen If it does not appear check if a treatment head is connected 8...

Page 40: ...the results are Bad the tuning has not succeeded A code representing the reason why the tuning has not succeeded occurs in the lower right corner see figure 12 Turn back the intensity knob to its zer...

Page 41: ...r_I2C_WriteErr I C Error during write User interface 20 err_LCD General LCD error 21 err_LCD_TimeOut LCD time out error 27 err_Ser_TimeOut Serial port time out E PROM errors 30 err_E2PROM General E PR...

Page 42: ...f the microcontroller Check the microcontroller and the software 00 xx If value differs from 00 an incorrect internal register bank is addressed Six short beeps will sound 01 err_IntRAM An error is fo...

Page 43: ...imer0 Timing of timer 0 is incorrect Check the microcontroller and the software xx Value gives an estimate of the number of times timer 0 has caused an interrupt since a 10 msec flag was set Six short...

Page 44: ...Mask for detecting an LCD error Only for internal use 31 err_E2PROM_Check sum Incorrect E PROM checksum Probably the unit is switched off during a write to the E PROM The checksum will be corrected a...

Page 45: ...ions between main PCB and ultrasound unit Check the power supplies Check the PWM xx Value represents the measured voltage Hexadecimal representation of the measured voltage 10 so at 22 5V decimal 225...

Page 46: ...err_VIsns Isns is too high if no ultrasound is produced V 0 05V Check the power stage in the ultrasound unit xx Value is the 8 bit measured voltage of Vusi so 0 05V 02h 8 Phase locked loop 60 err_PLL_...

Reviews: