USAGE INSTRUCTIONS
Press the handle (A) till the very end, and then insert the
spoke between the two pins on the tool (B). Make sure,
the spoke is completely parallel with the tool, which is
achieved by positioning the spoke to the side spacers (C)
on the two pins.
Do not release the handle, but gently release the pressure
and follow the path of the handle till it stops slowly,
like shown on the picture 2. If not certain, repeat the
measurement several times until you receive repeatable
results.
If you are uncertain of the spoke diameter, use the gauge
on the top of the tool (D).
Having the spoke diameter and the measured value,
check conversion table to determine the current tension.
*
We suggest to measure the spoke tension at normal room
temperature.
ENG
CALIBRATION INSTRUCTIONS
Insert the calibration spoke (E) into the Spoke Tension
Meter and use the same procedure as with a normal
spoke, like shown in the picture 4. After releasing the
handle, the indicator must show equal value to the one
specified on the calibration spoke.
If the value is not correct, screw the calibration bolt (F) on
the back side of the tool. If the indicator shows less that
it should, unscrew the calibration bolt counter clockwise
direction for a few turns. If the indicator is showing more,
screw the calibration bolt clockwise direction. Repeat the
process until you reach the desired value.
Calibration spoke must be stored in the box and only used
for calibration of this tool.
For additional help, contact Unior’s technical assistance at