Operation
79
Main menu - Options - System - General
The “Change signals after holding” command causes a smooth transition after
calibration or removal of the signal holding state. A sudden rise or fall of the
measured value in the analog output signal is avoided. The continuous transition
is specified in seconds.
General system values are configured in this menu.
The following must be observed here:
Change signals after holding:
When the Hold signals (mA) function is ended, an adjustment of the old and new
measured values is made over a time ramp.
Flushing time after cooling air or device start:
Specifies the length of time after the device start that the solenoid valve is open
until ignition starts.
Screen switch-off time:
Specifies the time after which the screen will be switched off if no input is made.
Display rate:
Refers to a time constant for various burners. It is preset by the manufacturer and
dependent on the burner type.
Carrier gas test cycles:
The time intervals for carrier gas calibrations, in hours, is defined.
Summary of Contents for CWD3000 EXP
Page 6: ...Technical data 6 1 2 Dimensions Fig 1 2 Housing dimensions Weight approx 140 kg ...
Page 14: ...EU Declaration of Conformity 14 ...
Page 30: ...Description and connections 30 ...
Page 56: ...Description of the work stations operator control elements 56 ...
Page 62: ...Operation 62 ...
Page 98: ...Decommissioning Switching off 98 ...
Page 103: ...Maintenance 103 ...
Page 104: ...Maintenance 104 ...
Page 110: ...Troubleshooting 110 ...
Page 112: ...Service 112 ...
Page 114: ...Disposal 114 ...
Page 116: ...Spare parts 116 ...