VOLT e SOCKET TEST
Fig. 1
Fig. 2
Fig. 3
Legend:
1.
Antiscratch glass
2.
Display TFT 1,77”
3.
Function
4.
TEST
5.
subfeatures selection
6.
subfeatures selection
7.
SCHUKO plug
This feature measure Voltage between phase and neutral and test the socket wiring
in according to local requirement.
Operation
Plug the instruments in the socket and select Voltage
function with button.Voltage between Phase and
Neutral will be shown at display if socket is correctly
wired and voltage contact is <50V (Fig 1) error due
to the uncorrect wiring is shown(Fig 2 e 3). Green bar
indicate every parameter is correct. Frequency can be
shown through button .
Wiring Error ( bar is
Red
and
sound allarm)
•
Phase - Neutral Reverse
•
Phase - Ground Reverse
•
Open Neutral
•
No Ground
•
Voltage contact >50V
Solution
Adjust the socket wiring
Adjust the socket wiring
Connect Neutral wire
Check Earth connection
Dangerous Voltage on mass