background image

18

17

7.USB Communication

1. Download and install the PC software (refer to UNI-T Documents Download 
    Operation Guide).
2. Connect the USB cable to the PC.  
3. Users can browse pictures and analyze data through the PC software. 
    Regarding its usage, retrieve the Software User Manual from the Help
    option of the operation interface.         

8. Specifications

Sensor

Temperature range

Measurement resolution

Mode

Accuracy

Measuring distance

Response time

Emissivity

IR resolution 

Pixel size

Color palette

Wavelength range

Field of view (FOV)

Spatial resolution (IFOV)

Thermal sensitivity (NETD)

Frame rate

Image format

UFPA

-10°C~400°C

0.1°C

Auto gain

±2.0°C or ±2%

25cm

500ms (95% of reading)

0.01~0.99 adjustable (default: 0.95)

19,200 pixels (160 × 120)

17μm

Iron red, rainbow, grayscale, red-white, blue-white

8μm~14μm

56° (H) × 42° (V)

11mrad

<50mK

<9Hz 

BMP

General parameters

PC software

Data transmission

Product size (L x W x H)

Display type

Display resolution

Battery

Auto power off

Battery life

Charging time

Charging voltage/current

Image storage

Drop test

IP rating

Certification

Storage temperature

Operating temperature

Operating humidity

Altitude 

Standard accessories

Yes

Type-C USB interface  

236mm x 75.5mm x 86mm

2.8” TFT LCD

320 × 240 pixels 

3.7V/5000mAh rechargeable Li-ion battery

5 minutes, 10 minutes, 30 minutes, 
off (default: 30 minutes)

6 hours

4 hours

5V/2A

Micro SD card

2m

IP65

CE

-20°C~60°C (-4°F~140°F)

0°C~50°C (32°F~122°F)

<90%RH (non-condensing)

2000m

User manual, USB cable, 16GB Micro SD card

9.Maintenance

Use a wet cloth or weak soap solution to clean the outer shell of the device.
Do not use abrasives, isopropyl alcohol or solvents to clean the outer shell, 
lens or window.

UTi165A User Manual

UTi165A User Manual

Summary of Contents for UTi165A

Page 1: ...UTi165A Professional Thermal Imager P N 110401108897X ...

Page 2: ...rantees that the product is free from any defect in material and workmanship within one year from the purchase date This warranty does not apply to damages caused by accident negligence misuse modification contamination or improper handling The dealer shall not be entitled to give any other warranty on behalf of Uni Trend If you need warranty service within the warranty period please contact your ...

Page 3: ... if it is damaged or modified to avoid inaccurate measurement results Please use the correct emissivity to obtain accurate temperature readouts To ensure the accuracy of the product please warm it up for 10 minutes before measuring if it has not been used for a long time When being charged the internal temperature of the product will rise which will lead to inaccurate temperature measurement So it...

Page 4: ...n Up button SET button 3 Display Display size 2 8 Display resolution 320 vertical x 240 horizontal pixels 1 2 3 4 5 6 7 8 9 10 11 12 13 Description Temperature unit option Cursor option Color palette option High Low temperature alarm option Settings option Minimum temperature Maximum temperature Description Minimum temperature point Maximum temperature point Center point Center point temperature D...

Page 5: ... Press the up down button to select C or F 5 Press the back button to exit the current menu Main menu Submenu Temperature unit option Cursor option Color palette option High Low temperature alarm option Settings option 6 Operating Instructions 6 1 Temperature Unit Selection 5 Menu Introduction 6 2 Center Point and High Low Temperature Tracking UTi165A User Manual UTi165A User Manual ...

Page 6: ...de them 7 Press the back button to exit the current menu 6 3 Color Palette Selection 1 Press the SET button to open the main menu 2 Press the left right button to select the option 3 Press the SET button to enter the color palette submenu 4 Press the up down button to select the desired color from Iron Red Rainbow Grayscale Red White and Blue White 5 Press the back button to exit the current menu ...

Page 7: ...u 2 Press the SET button to enter the Date Time submenu 3 Press the left right button to select the parameter to be adjusted 4 Press the SET button to enter the parameter adjustment state 5 Press the up down button to increase or decrease the value 6 Press the SET button to save the settings and return to set other parameters 7 Press the back button to exit the current menu 6 5 3 Emissivity To set...

Page 8: ...ress the back button to exit the current menu 6 5 6 Temperature Bar To turn on off temperature bar 1 Select the Temp Bar option up down button in the settings menu 2 Press the SET button to enter the Temp Bar submenu 3 Press the up down button to select On or Off 4 Press the SET button to confirm 5 Press the back button to exit the current menu 6 5 7 Alarm Temperatures To set alarm temperatures 1 ...

Page 9: ...to confirm 5 Press the back button to exit the current menu 6 5 10 Format SD Card To format SD card 1 Select the Format SD option up down button in the settings menu 2 Press the SET button to enter the Format SD submenu 3 Press the up down button to select Yes 4 Press the SET button to confirm 5 Press the back button to exit the current menu 6 5 11 Auto Save To turn on off auto save 1 Select the A...

Page 10: ...w grayscale red white blue white 8μm 14μm 56 H 42 V 11mrad 50mK 9Hz BMP General parameters PC software Data transmission Product size L x W x H Display type Display resolution Battery Auto power off Battery life Charging time Charging voltage current Image storage Drop test IP rating Certification Storage temperature Operating temperature Operating humidity Altitude Standard accessories Yes Type C...

Page 11: ...al Emissivity 0 85 0 96 0 75 0 14 0 96 0 09 0 06 0 95 0 98 0 96 0 93 Material Black paper Polycarbonate Concrete Copper oxide Cast iron Rust Gypsum Paint Rubber Soil Emissivity 0 86 0 8 0 97 0 78 0 81 0 8 0 75 0 9 0 95 0 93 Note This manual can be downloaded from the official website of Uni Trend refer to UNI T Documents Download Operation Guide Emissivity of common objects ...

Reviews: