UTG1000A Series Function Generator User Manual
74
4.2.3 Sweep Mode
Linear sweep: waveform generator changes output frequency in the linear way during the sweep;
Logarithmic sweep: waveform generator changes output frequency in logarithmic way; External sweep,
the default is linear sweep way, if need to change, please press Type Logarithm softkey.
4.2.4 Sweep Time
Set the required time from initial frequency to terminal frequency, the default is 1s, and the settable
range is from 1ms to 500s. If need to change, press Parameter →Sweep Time softkey in turn, then
enter number with number keyboard, and press corresponding unit softkey
4.2.5 Trigger Source Selection
When signal generator receives a trigger signal, it generates a sweep output, and then waits for the
next trigger signal. Sweep source can be internal, external or manual trigger. If need to change, press
Parameter →Trigger Source softkey in turn.
1)
When internal trigger is selected, waveform generator will output a continuous sweep, and the rate