Measurement Operation (2)
l
When AC voltage measurement has been
completed, disconnect the connection between the
testing leads and the circuit under test.
DC Voltage Measurement
(see figure 4)
The DC Voltage ranges are: 340.0mV, 3.400V, 34.00V,
340.0V and 1000V. To measure DC voltage, connect
the Meter as follows:
1. Insert the red test lead into the
HzV
terminal and
the black test lead into the
COM
terminal.
2. Set the rotary switch to
V
Hz
; DC measurement
is default or press
BLUE
button to select DC
measurement mode.
3. Connect the test leads across with the object being
measured.
The measured value shows on the display.
Note
l
In each range, the Meter has an input impedance
of 10M
. This loading effect can cause
measurement errors in high impedance circuits. If
the circuit impedance is less than or equal to 10k
,
the error is negligible (0.1% or less).
l
When DC voltage measurement has been
completed, disconnect the connection between the
testing leads and the circuit under test.
(figure 4)
black
red
Model UT60D: OPERATING MANUAL
18