7.2 Buttons (Picture 2, Picture 3)
Switch to insulation measurement. The tester defaults to insulation measurement after
booting. This button also can be used to exit each setting interface.
Short press to turn on/off timing measurement. The tester defaults
to continuous measurement mode after booting.
Short press to turn on/off comparative measurement. The tester defaults
to continuous measurement mode after booting.
Switch to voltage measurement (automatic identification of AC/DC voltage)
Short press the SET button to set system functions (short press the INSUL button to exit
each setting interface):
1. Date/Time
Set or adjust the system clock to match the current time and date.
2. Timing Measurement
Set the insulation measurement countdown time (from 60s). Once the time is up, the
insulation measurement will be stopped automatically.
3. DAR Measurement
Set the DAR measurement Time 1 (range: 15s~299s; 15s by default); set the DAR
measurement Time 2 (range: 60s~599s; 60s by default).
4. PI Measurement
Set the PI measurement Time 1 (range: 1min~29min; 1min by default); set the PI
measurement Time 2 (range: 10min~59min; 10min by default).
5. Factory Reset
Restore all settings data to default values.
When modifying a parameter, move the cursor with the LEFT/RIGHT button, use the
UP/DOWN button to change the current value, select the OK or Cancel option, and
short press the OK button to confirm the modification.
Short press to turn on/off USB transmission function. When the function is on, the
LCD displays the “USB” symbol.