Measurement Operation(12)
Note
l
The Meter automatically displays the temperature
value inside the Meter when there is no temperature
probe connection.
l
The included temperature probe can only be measured
up to 250. For any mesaurement higher than that,
the rod type temperature probe must be used instead.
l
When temperature measeuremnet has been completed,
disconnect the connection between the testing leads
and te circuit under test.
I. Capacitance Measurement
(see figure 9)
( figure 9)
Warning
To avoid damage to the Meter or to the equipment
under test, disconnect circuit power and discharge
all high-voltage capacitors before measuring capacitance.
Use the DC voltage function to confirm that the
capacitor is discharged. Never attempt to input over
60V in DC or 3V rms in AC to avoid personal dangerous.
Capacitance measurement has 4 measurement positions
on the rotary switch : 2nF, 20nF, 2
µ
F and 20
µ
F.
red
black
Model UT50E: OPERATING MANUAL