background image

This operating manual is subject to change without notice.

**END**

VI .Using the Meter

VII. Maintenance 

●  Turn on/off Power Save Mode

1. Press yellow button to turn on the Meter;

2. When the Meter turns on, it defaults at factory setup.

3. With the meter already turned on, pressing yellow button again will turn off the 

    Meter (UT331);

4. with the meter already turned on, press yellow button again will go into power 

    save mode (UT332, “     ” shows on LCD upper left corner). Repress to exit the 

    mode.

5. Set the sleep mode, the Meter will enter power save mode if no button is active 

    within the set time.

6. To ensure an accurate result, the user is recommended to wait for a while 

    before taking the readings if the measured humidity &temperature change 

    dramatically. 

● Data Hold

   Press HOLD button to freeze the reading, “HOLD” appears on LCD. Press again

   to quit the function.

● Turn on/off Backlight

 

   Under normal measurement status, press  “     ” button to turn on the backlight. 

   Press again to turn it off.(backlight is off under power save mode).

● View Max/Min Reading

   Press MAX/MIN button to check Max. and Min. readings. Long press the button 

   to exit the mode.

● View Dew Point Temperature

   Press ENTER to check dew point temperature, “DEW POINT” icon indicated on 

   LCD. Press again to quit the function.

● Store and Recall Data

   Press STORE button, the Meter will save the data into its memory and store up 

   to 100 sets of data (00--99). 

   The recorded data are those shown on the primary display. Press the button 

   again to exit the function.

   Press RECALL to review the data stored in the memory. Press again to exit the

   function.

1. Store Data

   Press STORE button and “DATA” icon shows on upper part of LCD. There are 

   two ways to store data.

   Manual Method: Preset time interval for auto storage at 00:00(refer to “Setting 

   Time Interval” section), then get into STORE mode, press ENTER once to save 

   current reading into current address of the memory, press ENTER again to save

   into next address. Press ▼▲ to change the address to store. “----” shows for 

   blank address.

   Auto Method: Preset time interval not at 00:00, then get into STORE mode, 

   press ENTER once, “:” icon flashes and the Meter will save the data in order into 

   the memory by beginning with current address and storing at a interval 

   previously set. Press ENTER again to pause the operation and “:” stops flashing. 

   Repress ENTER to continue.

2. Recall Data

   Press RECALL to check recorded data, “DATA” icon flashes. Press▼▲ to scroll 

   through recorded readings. The Meter defaults at the reading saved last time 

   you exit the STORE mode. Pressing ENTER under RECALL mode can quickly 

   review from No.00 to 99 data and also allow you to transfer data to PC through 

   USB. After quick review, the number stops at 99.

3. Clear Data

   Press STORE to go into STORE mode, “DATA” icon shows. Then long press 

   STORE for 2 seconds and “CLR” icon appears on the LCD, pressing ENTER 

   will clear all 100sets of data. Clearing item by item is not allowed. 

   “- - - -” shows for empty memory.

   Note: All 100 sets of data will be completely cleared once the function is enabled. 

   Step-by-Step clearing is not allowed. To avoid any data loss, please think it twice 

   before you use the function.

● Transfer Data

   The Meter will transfer data to PC under following two circumstances:
1. Enter into RECALL status, pressing ENTER button will not only allow you to 

    review the data quickly but also transfer the data automatically to PC.

2. When PC is running the supplied software and has been already connected to 

    the Meter, the measured data will be automatically transferred to PC.

Figure 3

● Cleaning the Casing

When it becomes necessary to clean the surface of the Meter, clean it with soft 

cloth or sponge dampened with just a little of water, soap water or mild detergent. 

Prohibit washing the Meter with water directly, which may damage the circuits

inside.

VIII. CE Certification

The Hygro-Thermometer conforms to the following standards.

● Replacing the Battery (See Figure 3)

When the Meter indicates “

    

”, please replace the battery timely. Follow the 

steps as below: 

EN61326: 2006

EN55022: 2006

EN55024: 1998+A1+A2

Function 

UT331 

UT332

Temperature:  -20  ~ 60
Humidity:        0%RH ~ 100%

Temperature:  -20  ~ 60
Humidity:         0%RH ~ 100%

Measuring 

range

Resolution

Measuring 

accuracy

Response 

time

Time
Data storage

Setup 

Functions

Temperature:  0.1
Humidity:        0.1%RH

Temperature:  0.1
Humidity:        0.1%RH

Temperature:  0  to 40

     

:

  0.5

                       -20  to 0    :  1.0

                       40  to 60   :  1.0

Humidity:

    

3.5%RH (10%RH to 90%RH)

                      

5.0%RH ( 10%RH or  90%RH)

Temperature:  0  to 40    : 

0.5

                      -20  to 0   : 

1.0

                      40  to 60   : 

1.0

Humidity:

    

2.5%RH (10%RH to 90%RH)

                      

5.0%RH ( 10%RH or  90%RH)

About 10 seconds (25 ,1m/s air)

About 10 seconds (25 ,1m/s air)

Relative time

Real-time time

0 to 99

0 to 99

Temperature unit, auto-save time interval, humidity offset, temperature offset, dew-point offset,  

sleep time setup and system time.

IX. Technical Specifications

7. Setting System Time

  Press SETUP to switch to system time mode, “S-t” icon shows. Press ▼▲ to  

  set up system time. Long press ▼▲ to increase the time quickly; Press ENTER 

  to select time format “h:m” or “m:s” . To make it easy to user, “m:s” format will 

  be automatically converted into “h:m” after the system time exceeds 60 seconds.

1. Unscrew the battery and remove it from the meter.

2. Replace the batteries with new 4×AAA batteries;

3. Rejoin the battery cover and tight up the screw.

P/N:110401104461X  DATE:2018.06.26  REV.5

Reviews: