Troubleshooting 

(1)

 

When press the button, the oscilloscope is blank screen. 

a. 

Check whether the power is correctly connected, power supply is normal or not. 

b. 

Check whether the power switch is really turned on. The soft power switch button should light in 
green when the instrument is normal boot. And it should have relay sound. 

c. 

If there is a relay sound, it indicates that the oscilloscope starts normally.   
Try the following steps: press the DEFAULT key, then press F1, if device returns to normal, it means 
backlight brightness is set too low.   

d. 

After completing the above steps, restart the oscilloscope. 

e. 

If the instrument still cannot work, please contact UNI-T. 

(2)

 

After signal acquisition, signal of waveform does not appear on the screen. 
a.

 

Check whether BNC port is connected properly.   

b.

 

Check whether the signal output channel has turned on. 

c.

 

Check whether the input signal channel has turned on. 

d.

 

Check whether has DC displacement in signal of information source. 

e.

 

Plug out the input signal, check whether the baseline is in the screen range (if not, please perform 
auto-calibration.) 

f.

 

If the instrument still cannot work, please contact UNI-T. 

(3)

 

The measured voltage amplitude value is 10 times larger or smaller than the actual value: Check the 
channel probe attenuation coefficient setting is consistent with the used probe attenuation rate.   

(4)

 

There is a waveform display but not stable 
a.

 

Check the trigger source in the trigger menu and confirm that it matches the input channel of the 
actual signal.   

b.

 

Check the trigger type: general signal should use “Edge” trigger. The waveform can display stably 
only the proper trigger type is used. 

c.

 

Change the trigger coupling setting to high-frequency or low-frequency reject, in order to filter out 
the high-frequency or low-frequency noise that interferes with the trigger. 

(5)

 

No display after pressing Run/Stop button 
a.

 

Check whether the mode at the trigger panel (TRIGGER) is on “Normal” or “Single” and whether the 
trigger level exceeds the waveform range.   

b.

 

If it exceeds

set the trigger level to the middle or set the trigger mode to Auto. 

c.

 

Press AUTO button could automatically finish the above setting. 

(6)

 

Waveform refresh is very slow: 
a.

 

Check whether the acquisition method is average and the average times are large. 

b.

 

Check whether the storage depth is the maximum. 

c.

 

Check whether the trigger release time is large. 

d.

 

Check whether the trigger is normal and the current time base is slow. 

e.

 

All the preceding causes slow refreshing of waveforms. Users can restore the factory settings so 
that the waveforms can be refreshed normally. 

Summary of Contents for MSO2000 Series

Page 1: ...Quick Start Guide MSO UPO2000 Series Digital Phosphor Oscilloscope...

Page 2: ...V1 00 Statement UNI T products are protected by patent rights in China and foreign countries including issued and pending patents UNI T reserves the rights to any product specification and pricing cha...

Page 3: ...n the Warning statement are not properly executed or observed Do not proceed to the next step until you fully understand and meet the conditions stated in the Warning statement Caution Caution indicat...

Page 4: ...from CAT III circuit or sockets more than 20 meters away from CAT IV circuit CAT III Primary circuit of large equipment directly connected to the distribution board and circuit between the distributi...

Page 5: ...s product before it is powered on AC power supply Please use the AC power supply specified for this device Please use the power cord approved by your country and confirm that the insulation layer is n...

Page 6: ...f UNI T for testing Any maintenance adjustment or parts replacement must be done by the relevant personnel of UNI T Cooling Do not block the ventilation holes at the side and back of this device Do no...

Page 7: ...Series digital phosphor oscilloscope includes four models Model Analog Channel Analog bandwidth LA AWG UPO2102 2 100MHz UPO2104 4 100MHz UPO2202 2 200MHz UPO2204 4 200MHz MSO2102 2 100MHz MSO2104 4 1...

Page 8: ...lease keep the packaging and notify both the transportation department and UNI T distributors UNI T will arrange maintenance or replacement 3 2 Before Use To perform a quick verification of the instru...

Page 9: ...and connect the probe to CH1 Make sure the probe s connector is properly connected with the oscilloscope Connect the probe s main clip and ground clip to the oscilloscope s compensating signal and gro...

Page 10: ...g waveform 14 Horizontal control HORIZONTAL 4 Jog dial knob 15 Analog channel input port 5 Function menu 16 Vertical control VERTICAL 6 Numeric keypad 17 Menu control soft key 7 Automatic control key...

Page 11: ...trend com 5 Rear Panel Figure 4 Rear Panel Table 2 Rear Panel No Description No Description 1 EXT Trig port 5 LAN 2 AUX Out and arbitrary waveform signal output 6 AC Power Input Socket 3 VIDEO Out 7 P...

Page 12: ...status identification 9 System time 2 Time base scale 10 USB DEVICE icon 3 Sampling rate storage depth 11 CH4 state icon 4 Waveform indicator 12 CH3 state icon 5 Waveform trigger position 13 CH2 stat...

Page 13: ...on the screen and then to setup Tap the function guide icon on the left corner of the screen to enable it Tap to pop out numeric keypad to set parameter Tap virtual keyboard to set label and file nam...

Page 14: ...ng Rectangle Open function guide and tap drawing rectangle icon to switch drawing rectangle mode Drag finger on the screentodrawing rectangle asshowninFigure 10 a 10 b Move fingeraway fromthescreen me...

Page 15: ...5 Touch Shortcut Operating Separate two fingers and meanwhile slide two fingers to one direction this gesture can quickly turn on off time measurement and voltage measurement cursor a Time measuremen...

Page 16: ...This symbolrepresentsthat numeric keypad on the front panel or touchto pop out numeric keyboard to input content This symbol represents that the current page can be flipped up or down Circle numbers...

Page 17: ...red voltage amplitude value is 10 times larger or smaller than the actual value Check the channel probe attenuation coefficient setting is consistent with the used probe attenuation rate 4 There is a...

Page 18: ...5 30pm UTC 8 Monday to Friday or via email Our email address is infosh uni trend com cn For product support outside mainland China please contact your local UNI T distributor or sales center Many UNI...

Reviews: