21
Environmental temperature fluctuates.
Wait for the environment temperature to
stabilize. It is influenced by the
principle of the Pirani gauge.
Hot or cold gas is flowing.
Please do not be affected by the gas
temperature. It is influenced by the
principle of the Pirani gauge.
Effects of moisture and dirt.
Please do not let it affect you as much as
possible. Evaporation heat is affected by
gradual evaporation.
There is a leak in the sensor part or near
where the vacuum gauge is attached.
Sensor part leak
⇒
Replace the vacuum gauge.
Other leaks
⇒
Please stop the leaks.
Poor contact of USB cable, increase in
wire resistance, etc.
Check the connection of the USB cable,
etc.
USB cable is electromagnetically induced
(due to external noise)
Change the installation location of the
USB cable, or turn off the equipment that
may be a source of noise.
●
Expected pressure and measured pressure differ greatly
Cause
Action
Contamination inside the sensor part.
Sensitivity deviates significantly.
Check with another vacuum gauge. If the
other vacuum gauges are fine, then the
vacuum gauge is the problem.
You are measuring the air on the nitrogen
setting.
Normal.
When measuring air in the nitrogen
setting, the gas dependence of the Pirani
gauge will indicate higher pressures.
An environment with a lot of moisture and
oil.
Install so that there is no influence as
much as possible. Evaporation heat is
affected by gradual evaporation.
The actual pressure is different.
Check with another vacuum gauge.
Summary of Contents for SWU10-U
Page 6: ...SK00 9662 EI 103 00 VI Revision history Date Number Reason 2022 12 14 00 First edition...
Page 34: ...SK00 9662 EI 103 00 27 9 EC DECLARATION OF CONFORMITY...
Page 35: ...SK00 9662 EI 103 00 28 10 UK DECLARATION OF CONFORMITY...
Page 36: ...SK00 9662 EI 103 00 29 11 Certificate of decontamination...
Page 37: ...SK00 9662 EI 103 00 30 12 Drawings 12 1 SWU10 U dimensions...
Page 38: ...SK00 9662 EI 103 00 31 12 2 SWU10 U dimensions...