Option 

1.3.1.

 

Sensor head 

Sensor head 

WP-01/WP-03/WP-03/WP-16 

WPB-10-034 

Filament material 

Pt 

Pt 

Other materials 

BS/Ni-plating 

Ni

Kovar

Glass

SnSbCu 

SUS304

Cu(Gasket) 

FeNiCo(Ni-plating)

Al

2

O

3

 

Permitted leak ratio 

1

×

10

-8

Pa

m

3

/s  max 

1

×

10

-8

Pa

m

3

/s  max 

Weight(g) 

WP-01

about  26

WP-02

about  45 

WP-03

about  63

WP-16

about  77 

about  173 

Internal volume(cm

3

WP-01

about  19

WP-02

about  17 

WP-03

about  17

WP-16

about  22 

about  18 

Pressure max

<  2

×

10

+5

Pa(abs) 

<  2

×

10

+5

Pa(abs) 

Bakeout

80

  max 

250

  max 

1:The breakdown pressure of the flange and clamp are to be considered otherwise. 

2:Bakeout temperature is a temperature of the sensor unit.  Please remove the electronics(GP-H)  or  the 

conversion cable(GP-BH). 

3:WP-04 andWP-05  discontinued producing in August, 2012. 

 

1.3.2.

 

Measuring unit 

Measuring unit 

GP-H 

GP-BH

Conversion cable 2m

 

Sensor head 

WP-01

WP-02 

WP-03

WP-16 

WPB-10-034 

Weight 

35g 

GP-BH: 0.13kg 

Conversion calbe: 0.2kg 

remarks 

GP-H for WP-01/WP-02  and   

WP-03/WP-16  are different. 

 

 

1.3.3.

 

Sensor head cable 

Sensor head cable 

2

5

10

15

20

30

50

100m 

Weight(kg) 

0.2/0.4/0.7/1.0/1.3/1.9/3.9/7.7 

 

1.3.4.

 

Supply cable 3m

only GP-2001G

 

The plug specification of supply cable is AC125V / 7A. 

 

1.3.5.

 

Others 

JCSS calibration certificate 

Only combination with sensor unit 

General proofreading test 

Only combination with sensor unit 

Inspection certificate 

 

Calibration certificate 

 

 

 

This gauge is based on the measurement of thermal conductance and the 

indicated value may change depending on the type of the sensor head. Some 
sensor heads may require change of the circuitry or structure of the measuring 
unit. Therefore, if the type of the sensor head is changed after delivery of the 
gauge, re-adjustment will be required. Factory adjusted conditions (type of 
sensor head) are inscribed on the measuring unit GP-H. Check them before 
operation. 

 

Summary of Contents for GP-2001G

Page 1: ...nual is applicable to the following gauges Model Serial No GP 2001G 0001G and higher GP 2002G 0001G and higher Components Division ULVAC Inc 2500 Hagisono Chigasaki City Kanagawa ken Japan http www ul...

Page 2: ......

Page 3: ...ety Symbols Safety symbols are used throughout this instruction manual to call the operator s attention to safety The terminology used in safety symbols is classified below Indicate status of urgency...

Page 4: ...will be damaged or it will pop out from the connector causing injury to the surrounding including human body If the pressure exceeds atmospheric pressure provide an isolation valve so that the pressur...

Page 5: ...d in an atmosphere that can cause hazards to the human body must be disposed of by a specialist in disposal Maintenance The electrical circuitry inside this unit uses aluminum electrolytic capacitors...

Page 6: ...head installation method to GP H Section 4 1 Changed the overall operation diagram Chapter 6 Change fuse specification Chapter 7 Change warranty Chapter 8 Added certificate of Decontamination Nov 19...

Page 7: ...ing the supply cable 7 2 2 4 Electrical connection 7 2 2 5 A sensor head installation method to GP H 9 3 OPERATING PROCEDURE 10 3 1 CAUTIONS IN HANDLING 10 4 OPERATING PROCEDURE 11 4 1 SUMMARY OF THE...

Page 8: ...rani vacuum gauge Measurable pressure range 0 4 to 3000 Pa Measuring point 1 point Display 7 segment digital Indicator red LED 0 0 3000 0 Pa Display sampling time 100 msec Measurement accuracy 4 0 10...

Page 9: ...on cable 2m Sensor head WP 01 WP 02 WP 03 WP 16 WPB 10 034 Weight 35g GP BH 0 13kg Conversion calbe 0 2kg remarks GP H for WP 01 WP 02 and WP 03 WP 16 are different 1 3 3 Sensor head cable Sensor head...

Page 10: ...section 4 for the operating procedure 2 Comparison setpoint output action display The green LED comes on when the comparison output is ON 3 PV indicator Main indicator that displays the comparison set...

Page 11: ...orrecting function input comparison setpoint output recorder output etc A 3mm dia solderless terminal is best suited to connection When connecting the input output wiring do not make mistake in wiring...

Page 12: ...ike metal fragments or combustibles are admitted into the vacuum gauge through the air vents or other openings remove them Otherwise the vacuum gauge may be damaged Operating environment Do not use th...

Page 13: ...sure that the power supply cable crimping terminal section does not touch other conductive components on the control etc Check power supply voltage Before turning on the power supply check that the va...

Page 14: ...connection For safety reasons ground the GND terminal of the controller Connect the power cable to the connecting terminal block No 19 and 20 on the rear panel using a solderless terminal Before conne...

Page 15: ...8 Fig 2 3 GP 200 G overall connection diagram CONVERSION CABLE WBP 10 GP 200 G SENSEOR HEAD CABLE GP H BH...

Page 16: ...GP H to a sensor head By the kind of the sensor head installation met hods are different For sensor head WP 01 02 For sensor head WP 03 16 It is inserted a screw for fixation by the vent sid e of the...

Page 17: ...or head ON during the measurements 3 If the sensor head is exposed to chemically active gas or highly adsorbing gas its characteristics may change In such a case repeat pumping flushing operation afte...

Page 18: ...V indicator 0 Lin0 SV indicator 1 Lin1 SV indicator 2 Lin2 SET Pressing SV indicator SP1 PV indicator setpoint value SV indicator SP2 PV indicator setpoint value SV indicator SP3 PV indicator setpoint...

Page 19: ...LAY SP1 DISPLAY SP1 SP2 SP3 SP1 ALM SP2 ALM SP3 ALM SET SET SET SP2 DISPLAY SP3 DISPLAY SP1 SETTING SP3 SETTING SP2 SETTING CANCEL CANCEL CANCEL FINALIZE FINALIZE FINALIZE 1 The comparison value setti...

Page 20: ...AY Data can be set with the switch and of the data can be selected with the switch If is displayed in the portion on the SV indicator shown in the figure above each press on the switch decrements the...

Page 21: ...value are not correlated with each other refer to 4 5 Adjusting the Output Voltage The ripple of the recorder output is approx 10 mV p p Lin 0 Lin 1 Lin 2 Pressure Pa Output Voltage V Pressure Pa Outp...

Page 22: ...n1 Output Fig 4 3 Lin2 Output 0 1 2 3 4 5 6 7 8 9 10 11 12 0 200 400 600 800 1000 1200 1400 1600 Recoder Output V Pressure Pa 0 1 2 3 4 5 6 7 8 9 10 11 12 0 20 40 60 80 100 120 140 160 Recoder Output...

Page 23: ...ary to make zero resetting at a pressure higher than 1 0 Pa replace the sensor head recalibrate the vacuum gauge or take other actions because the sensor head contamination is responsible 1 Setting an...

Page 24: ...n A is a coefficient referred to as free molecule thermal conductivity m3 K 1 and its value varies with the type of gas is a coefficient that corrects the imperfectness of exchange of amount of energy...

Page 25: ...on C and is in inverse proportion to MT In a high pressure region the number of molecules that carry heat energy increases while the free path of molecule becomes short so that the amount of transport...

Page 26: ...E 1 1 0 1 8 4 3 8 0 1 7E 1 5 1E 1 1 1E 2 2 3E 2 8 5E 2 8 0E 3 5 9E 1 1 2 2 3 5 3 1 1E 1 2 1E 1 5 2E 1 1 1E 2 2 1E 2 5 3E 2 1 0E 3 3 7E 1 8 2E 1 1 7 4 7 9 8 2 1E 1 4 0E 1 9 0E 1 2 4E 2 2 0E 3 7 4E 1 1...

Page 27: ...20 Change of Indication with Ambient Temperature Fig 5 2 Change of Indications with Ambient Temperature 0 10 20 30 40 50 1 10 100 1k 10k Indication of Pirani gauge Pa Ambient Temperature...

Page 28: ...1 Normal Symptom The PV display remains and the SV display remains OFF Possible cause Corrective action The sensor head filament has burnt out 1 Check the filament for continuity referring to Fig 10 4...

Page 29: ...s been adjusted at 25 C 1 Change the place of installation Install in a clean well ventilated place not blown with draft Symptom Pressure display does not show a constant value Possible cause Correcti...

Page 30: ...local ULVAC representatives for repair Return charge shall be paid by Buyer Disclaimer 1 Failure occurred after expiration of warranty period 2 Failure caused by force majeure such as fire storm and f...

Page 31: ...t for names and contents of hazardous substances or elements Name of parts Hazardous substances or elements Pb Hg Cd Cr6 PBB PBDE Printed Circuit Board Chassis Connector AC DC Converter Label Detectio...

Page 32: ...25 9 CERTIFICATE OF DECONTAMINATION...

Page 33: ...26 10 RELATED DRAWINGS Fig 10 1 Dimensional drawing for GP 200 G 12 5 123 5 14 44 Panel hole Demention DATA ALM GP 2000G kPa Pa 91 5 44 5 91 48 96 SP1 SP2 SP3 SET ZERO...

Page 34: ...27 Fig 10 2 Dimensional drawing for measuring unit GP H Fig 10 3 Dimensional drawing for measuring unit GP BH...

Page 35: ...28 Fig 10 4 Sensor head filament connection diagram Pin No Use 1 5 V 2 15 V 3 V 4 V 5 AG Cover FG Fig 10 5 Sensor head cable 2 5 10 15 20 30 50 100m...

Page 36: ...ial Case material WP 01 18 Pt 25 BS Ni plating WP 02 15 18 Pt 25 BS Ni plating WP 03 3 8 Pt 25 BS Ni plating WP 16 NW 16 30 Pt 25 BS Ni plating WPB 10 034 UFC 034 Pt 25 SUS304 Fig 10 6 Sensor heads co...

Reviews: