84
22.4.7.
Reading the setpoint 3 value
Command
:
AD0
AD1
3R
CHKH
CHKL
CR
Format of returning from this instrument to PC
:
AD0
AD1
3
X
.
X
X
E
±
X
X
CHKH
CHKL
CR
* “
±
” denotes “+” or “-“.
22.4.8.
Writing the setpoint 1 set value
Command
:
AD0
AD1
1W
X .
X
X
E
±
X X
CHKH
CHKL CR
* A numeric value that can be set varies with each sensor. If a value lower than the lower limit
set value of each sensor is set, it will be set at the lower limit value of setting or, if a value
higher than the higher limit set value, it will be set at the higher limit value of setting.
* Internal mantissa part comparison processing is done at 2 digits below decimal point [
□
.
□
□
×
10
-
].
* If a command is normally received, ‘o’ of normal receipt will be returned.
22.4.9.
Writing the setpoint 2 set value
Command
:
AD0
AD1
2W
X
.
X
X
E
±
X
X
CHKH CHKL CR
* A numeric value that can be set varies with each sensor. If a value lower than the lower limit
set value of each sensor is set, it will be set at the lower limit value of setting or, if a value
higher than the higher limit set value, it will be set at the higher limit value of setting.
* Internal mantissa part comparison processing is done at 2 digits below decimal point [
□
.
□
□
×
10
-
].
* If the command is normally received, ‘o’ of normal receipt will be returned.
22.4.10.
Writing the setpoint 3 set value
Command
:
AD0
AD1
3W
X
. X
X
E
±
X
X CHKH CHKL CR
* A numeric value that can be set varies with each sensor. If a value lower than the lower limit
set value of each sensor is set, it will be set at the lower limit value of setting or, if a value
higher than the higher limit set value, it will be set at the higher limit value of setting.
* Internal mantissa part comparison processing is done at 2 digits below decimal point [
□
.
□
□
×
10
-
].
* If the command is normally received, ‘o’ of normal receipt will be returned.
22.4.11.
SW1 ZERO point adjustment command
Command
:
AD0
AD1
ZER
CHKH
CHKL
CR
* Adjustable pressure range: Approx.
±
1 Pa
* Signal for zero point adjustment is output from the I/O connector.
* If a command is normally received, normal receipt ‘o’ will be returned though adjustment is
not made. Read the measurement value and others after adjustment and check them.
22.4.12.
SW1 atmospheric pressure adjustment command
Command
:
AD0
AD1
ATM
CHKH
CHKL
CR
* The adjustable pressure range is 1
×
10
-4
to 2
×
10
-5
Pa
* Signal for atmospheric pressure adjustment is output from the I/O connector.
* If a command is normally received, normal receipt ‘o’ will be returned though adjustment is
not made. Read the measurement value and others after adjustment and check them.
22.4.13.
SW1 zero point, atmospheric pressure adjustment correction reset command
Command
:
AD0
AD1
CLR
CHKH
CHKL
CR
* Reset signal is output from the I/O connector.
* If the command is normally received, normal receipt ‘o’ will be returned.