Using the Ultrawave Cycle Validation Software
If the MC card Validation mode is selected, cycle information is saved to the inserted MC
card and data can be saved in an electronic format. The following pages explain how to
use the MC card and the Ultrawave Cycle Validation Software.
1.
Setting the Validation Mode
This is done in Screen 1 of the
Administrator Instruction
menu (see pg. 4). In order to
enable digital cycle validation set the Validation Mode to MC Card only or Printer & MC
Card.
2.
Inserting the MC Card
In order for data to save to the MC card, the card must be inserted into the port located
on the left hand side of the control panel before your
Hygea 6427
is switched on.
The digital cycle traceability will save the information of each cycle (as shown on the
ticket on page 23) to the MC card. This information can then be transferred to a PC and
an electronic copy saved, using the Ultrawave Cycle Validation software supplied on the
MC Card.
If the MC card is not inserted, then a message will display on the screen in Operator mode
“MC Card not present”. If MC Card only is selected as the Validation mode, and the MC
card not present, then the cycle will not run.
If no MC card is available, then the chosen Validation Mode will have to be amended in
the Administrator mode menu.
Insert MC
card here