Ultrawave Hygea 6427 Operator'S Instruction Manual Download Page 16

 

 

12 

On  successful  completion  of  the  cycle,  the  following  message  will  be  displayed  and  the 
data recorded. 

 

 

If  the  print  option  is  chosen,  then  at  the  end  of  each  cycle,  the  integral  printer  in  the 

Hygea 6427

 will print a ticket showing the details of that cycle: 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Once the ticket is printed, simply tear off and store as necessary. 

 

 

 

 

 

 

 

 

Ultrawave

 

 

Version: V3-00. 
Model: HYGEA 6427 

 

Serial- No: 056204 
Time: 15.30 

 

Date:  01/12/17 

Hops A 

 

Op: 01            Nurse A 
Cycle-No:       01234 
 
Set-Parameters: 
Min-Temp:         

20.C 

Set-Temp:         

25.C 

Max-Temp:        

40.C 

Cycle-Time:      

15.M 

 
Cycle-Parameters: 
Cycle-Start-Temp:        

25.C 

Min-Temp:          

25.C 

Max-Temp:         

26.C 

Cycle-Finish-Temp: 

26.C 

 
Cycle-Time:       

5.00M 

Ultrasonic-Action:     

PASS 

*Cycle PASS* 

 

Summary of Contents for Hygea 6427

Page 1: ...Hygea 6427 Operator Instruction Manual...

Page 2: ......

Page 3: ......

Page 4: ......

Page 5: ...intenance Cleaning 21 Periodic testing and validation 22 Validation Testing Reminder 23 Hygea 6427 Cycle validation 24 Using the Ultrawave Cycle Validation Software 25 Hygea 6427 good practice 30 Dete...

Page 6: ...so may damage the bath and invalidate your warranty Care should be taken when operating the bath at higher temperatures as external surfaces may become hot For best cleaning results use a detergent at...

Page 7: ...card 1 x Hygea 6427 Administrator instruction manual 1 x Hygea 6427 Operator instruction manual How to install The Hygea 6427 should be installed on a benchtop or stand if provided located close to a...

Page 8: ...ated to the minimum operating temperature the Hygea 6427 is ready to use although an extra degas cycle may be required Select your operator identification Press the ENTER key to start the cleaning cyc...

Page 9: ...stainless steel clips To use these press the small stainless clip in towards the pipe and pull the connector straight up the connector is then pushed into the other receptacle 1 The tubes can be conne...

Page 10: ...tor into the tube and then connecting the hollow instrument 5 The instrument and hose can then be connected to the manifold Important The tubes must be connected whenever the Hygea 6427 is in use not...

Page 11: ...first switched on the following screen will appear for approximately 5 seconds While this screen is displayed the buttons are inoperative This is normal as the Hygea 6427 is running through its start...

Page 12: ...wing screen will display The Hygea 6427 will automatically start heating and degassing the liquid When the liquid reaches the required temperature the following screen will be displayed and the Hygea...

Page 13: ...down the list indicating which name is selected Press the Next button to show the next group of four names Up to 20 names can be stored in the operator list When the correct operator is selected pres...

Page 14: ...ays to confirm that detergent has been added Press the Start button to confirm that the liquid contains detergent See page 30 for information about detergents The Hygea 6427 will then begin the cleani...

Page 15: ...d 11th minutes If the pump time is set to 2 minutes the pump will run during the 1st and 2nd 6th and 7th and 11th and 12th minutes While the pump is in operation the word pump will display on screen W...

Page 16: ...ng the details of that cycle Once the ticket is printed simply tear off and store as necessary Ultrawave Version V3 00 Model HYGEA 6427 Serial No 056204 Time 15 30 Date 01 12 17 Hops A Op 01 Nurse A C...

Page 17: ...on page 12 to the MC card This information can then be transferred to a PC and an electronic copy saved using the Ultrawave Cycle Validation software supplied on the MC Card It is recommended that th...

Page 18: ...e screen displays a Low Fluid warning as below the bath will not operate until the unit is filled to the correct level and the detergent dose added If the screen shows a Lid Open warning the cycle wil...

Page 19: ...ng press and hold the ENTER button at the bottom of the screen until the warning stops If the operator aborts the cleaning cycle the following screen will be displayed If the fluid level drops below t...

Page 20: ...d on the MC card when the machine is next switched back on The cycle will be recorded as an aborted cycle If the cycle fails due to a power failure the following screen is displayed when the power is...

Page 21: ...lowing screen will display If the ultrasonic power drops below a certain point during a cycle the cycle will stop and the following screen will be displayed If the cycle aborts the reason will be disp...

Page 22: ...sure that the drain is either directly over a sink or is connected to a tube which is placed into a sink Step 2 Turn the tap until open DRAIN TAP SHOWN OPEN Step 3 Allow the liquid to drain Step 4 Clo...

Page 23: ...new paper rolls need to be ordered Use the paper feed button to feed through the last few centimetres of paper and fit a new roll Step 1 Open the printer paper tray by pulling the clear cover towards...

Page 24: ...g Step 5 Close the cover Step 6 Close the printer paper tray Step 7 Feed a small amount of paper through the printer by pressing the paper advance button on the right hand side of the printer or the p...

Page 25: ...nfected daily Cleaning and disinfection processes may be regulated by the CSSD HSDU manager or AP so check with these before attempting manual disinfection Check hoses for any signs of visible blockag...

Page 26: ...available from Ultrawave If there is no access to the Activity Meter this test should be replaced by the aluminium foil ablation test 3 Quarterly tests carried out by the operator Estates department U...

Page 27: ...Therefore after 11 months in operation a reminder will appear on the LCD display warning that the annual validation testing is due The Validation Testing reminder will display as below When this remin...

Page 28: ...more information If the hard copy print out is chosen then at the end of each cycle the integral printer in the Hygea 6427 will print a ticket showing the details of that cycle Once the ticket is prin...

Page 29: ...rd the card must be inserted into the port located on the left hand side of the control panel before your Hygea 6427 is switched on The digital cycle traceability will save the information of each cyc...

Page 30: ...IT administrator Also ensure that Microsoft Net 3 5 Framework is installed as well as the associated updates In order to install the Software the MC Card should be inserted into the PC using the USB c...

Page 31: ...lready saved to the PC the data will display in the below format If you wish to copy new cycle data to the software click on the Copy Cycle Results from MC Card button You will need to ensure that the...

Page 32: ...by pressing the print button It is also possible to print data from multiple cycles This can be done by clicking on multiple cycles in the left hand column In order to print the data from multiple cyc...

Page 33: ...in the applicable search box located in the top right of the screen The following fields can be used to search for cycle data Serial Number of the machine Location Clinic Practice Operator name Date...

Page 34: ...ning medical instruments the maximum temperature should be set to 40 C Setting the Max above 40 C may cause proteins to coagulate and bake on to the instruments Ultrasonic activity itself will heat up...

Page 35: ...keep your bath clean Not only will contaminated liquid reduce the performance of the bath it may also damage it Change the cleaning liquid regularly Your cleaning process will determine how often to...

Page 36: ...g is now 80 decibels There is also an exposure limit value of 87 decibels taking account of any reduction in exposure provided by hearing protection above which workers must not be exposed To help you...

Page 37: ...EN 13202 2000 Ergonomics of the thermal environment Temperatures of touchable hot surfaces Guidance for establishing surface temperature limit values in production standards with the aid of EN 563 199...

Page 38: ...able acidic caustic or corrosive chemicals or fluids not recommended by Ultrawave The user should familiarise themselves with this instruction booklet before operating the equipment and should apply t...

Page 39: ...s message displays run another cycle to see if the problem persists The screen displays OVER TEMP Drain and refill with fresh cold water The Hygea 6427 is fitted with a temperature sensor that automat...

Page 40: ...tect the health and safety of Ultrawave employees reducing the risk of potential injury or infection More information on decontamination guidelines as well as copies of decontamination certificates ca...

Page 41: ...tor test strips and holders Brownes strips and holders for weekly testing of the Hygea 6427 Service Contracts and Validation Testing Ultrawave can offer full Service and Validation packages for the th...

Page 42: ..._______________ __________________________________________________________ __________________________________________________________ __________________________________________________________ _______...

Page 43: ...39...

Page 44: ......

Page 45: ......

Page 46: ...ce Tel 44 0 845 330 4238 Fax 44 0 845 330 4231 www ultrawave co uk admin ultrawave co uk service ultrawave co uk Ultrawave baths are manufactured in the United Kingdom Ultrawave baths are manufactured...

Reviews: