www.ultravation.com
5
UVCMONITOR (FOR ME SERIES ONLY)
The digital display shows the percent UVC output
based on a new UVC lamp. The output needs to be
calibrated after 100 hours of use. This allows the
lamp output to stabilize.
After 100 hours of operation the SCALE
adjustment – on the right side of the control box –
must be adjusted until the percent output = 100.
ANNUAL MAINTENANCE PROCEDURE:
Annual Germicidal Lamp Replacement and Quartz
Cleaning:
1. Unplug the power supply from surge protector and
the electrical outlet.
2. Pull the black cap from the center of the clear
gland nut.
3. Carefully lift the germicidal lamp out of the quartz
sleeve by pulling up on the gray harness wire and
discard the lamp. Remember the lamp is glass
and can cause injury to you if the lamp is
dropped.
4. Open the boiler drain on outlet side of UV unit to
depressurize vessel. When completed, close
boiler drain.
5. Carefully remove clear gland nut, gently grasp
quartz sleeve and lift out.
6. Remove clear gland nut and O-ring from quartz
sleeve.
7. Clean quartz sleeve with household alcohol. If
stained, replace.
(Reference Quartz Installation)
8. Follow with germicidal lamp installation.
9000 Hour Timer Reset (FOR SV SERIES ONLY):
It is recommended to reset the 9000 hour timer after
replacing the germicidal lamp.
1.
Make sure the electrical power cord is removed from
the electrical outlet.
2.
Remove power supply cover.
3.
The battery which operates the 9000 hour timer during
a power outage should be changed annually with a
CR2032 (3V) battery or equivalent. To remove old
battery, pull it up and away from the battery harness
located on the right side of circuit board.
When installing the new battery, make sure the “+” side of the
batter is viewable.
Locate the brown reset button on the lower center portion of
the green circuit board, to the left of the battery.
1.
Once located, hold reset button for 5 seconds and
release.
2.
Re-install the power supply cover and plug electrical
power cord into the dedicated outlet
.
UVLampMonitor (FOR SV SERIES ONLY):
LED
Color
Audible
Visual Alarm
Unit
Status
Green
No audible
OK
Solid Red
No Audible
Reset lamp service
indicator or replace lamp
within 30 days
Pulsing Red
(Every 5
seconds)
Beep every
60 Minutes
Lamp has exceeded 1
year of operation reset
or replace lamps
Flashing Red
(Every 1 second)
Beep every
1 second
UV lamp(s) are not lit