         Safety In Use

 

 

PFM1OpManualIssue5                                                                                                                     Page 

2

 

 
 

 

Special  Precautions

 

 

Your 

PFM1 Antenatal Cardiotocograph

 has been designed for 

electrical safety.  All the safety and operating instructions should be 
read before operating the

 PFM1

.  Failure to do so could result in 

injury to the user, patient, or damage to the system and accessories. 

 
Electrical Shock Hazard 
 

Do not defeat the grounding integrity of this system.  Protection against electrical shock, 
in the event of failure of basic insulation, is provided by the connection of the chassis to 
the safety ground.  Safety grounding occurs only when the 3-wire cable and plug provided 
with the system are connected to a properly grounded receptacle. 

 

Do not remove the system cover.  The system should be serviced by trained and qualified 
personnel only.  Contacting the hazardous voltages within the system could cause 
serious injury. 

 

Do not use the system if the power cord has any cuts or openings. 

 

Do not use the transducer if the cable has any cuts or openings. 

 

Do not use the transducer if the transducer face is cracked or chipped. 

 

Do not immerse the transducer cable connectors in any liquids.  

 

Should the electrical safety fuses have to be replaced, use only fuses of the same type 
and rating. 

Don not use this device at or close to water deliveries! 

 
 
Explosion Hazard 

 

Do not operate or use this system in the presence of flammable 

anesthetics or gases as it could lead to explosion. 

Summary of Contents for PFM1

Page 1: ...Operating Manual PFM1 Fetal Monitor Ultrasound Technologies Ltd Foresters House Itton Chepstow Gwent NP6 6BZ U K Telephone 44 0 1291 641301 Fax 44 0 1291 641302...

Page 2: ...Toco Zero 4 Indicators and Connectors 5 Power On Indication 5 RS232 Output 5 Remote Patient Event Marker 5 Probe Connection 5 Operating Procedures 6 User Configuration 6 The Printer 8 Ultrasound Feta...

Page 3: ...mydoppler de A service manual containing circuit descriptions diagrams parts and spares lists is available for the PFM1 by contacting the address above To maintain the PFM1 s performance it is recomm...

Page 4: ...ided with the system are connected to a properly grounded receptacle l Do not remove the system cover The system should be serviced by trained and qualified personnel only Contacting the hazardous vol...

Page 5: ...ansducer or cable damage could occur Symbols Used The following symbols are used on the PFM1 and are in accordance with BS EN60601 1 1990 Where they are associated with the connection of external equi...

Page 6: ...change user options Volume Down This button is used to decrease the audio volume level or in configuration mode to change user operations US1 US1 Transducer Select and Twins trace mode Pressing this...

Page 7: ...t can be applied to this output is 15VDC WARNING Any external equipment connected to this output must meet the relevant safety standard Remote Event Marker Connection of an external event mark switch...

Page 8: ...Up button One press advances the day by one pressing the Volume Up button for longer advances the day by 10 To change the year press the Volume Down button One press advances the year by one pressing...

Page 9: ...a Block is printed at the start of each recording this function can be disabled by setting Data Block to off It changes from on to off and back again by pressing the Volume Up button When the Data Blo...

Page 10: ...al side of the paper is uppermost Pull the paper out of the front of the unit making sure it is positioned below the door catches Close the door and latch by pressing down firmly on it To operate the...

Page 11: ...only showing an outline heart then this is an indication that the signal quality is not optimum Improved recordings can be obtained by repositioning the transducer so that this indicator always shows...

Page 12: ...the ingress of fluids NO coupling gel is required for this transducer to operate correctly Place the transducer on the centre line over the fundus in a position where the uterus is firm and secure in...

Page 13: ...he NST is summarised Reactive accelerations of 15 BPM or more which last for 15 seconds or longer twice in any 10 minute period These may be accompanied by fetal activity Non reactive evidence of feta...

Page 14: ...checking fuses and connections Check all connections and fuses For replacement fuses consult the technical specifications for specific ratings required No display information on LCD Verify the system...

Page 15: ...se on any damaged part contact them immediately Cleaning Enclosure Clean the exterior of the system with a soft dry cloth In the event of stubborn spots disconnect the system from the power source Use...

Page 16: ...Data Presentation Strip chart recorder and alphanumeric display module Printhead 4 inch thermal solid state printhead Resolution 8 dots mm Speeds 1 2 3 cm min Paper Roll or Z fold Paper type Plain bla...

Page 17: ...lving adverse EMC conditions Emissions Care has been taken through the design and manufacturing processes to minimise the EM emissions that may be produced by this equipment However in the unlikely ev...

Reviews: