Operating Manual 

 

PFM1 

 Fetal Monitor 

 
 
 
 
 
 

Ultrasound Technologies Ltd

      Foresters House, Itton, Chepstow, Gwent.  NP6 6BZ  U.K. 

Tel44 (0) 1291 641301        Fax +44 (0) 1291 641302

 

 

 

Summary of Contents for PFM1

Page 1: ...Operating Manual PFM1 Fetal Monitor Ultrasound Technologies Ltd Foresters House Itton Chepstow Gwent NP6 6BZ U K Telephone 44 0 1291 641301 Fax 44 0 1291 641302...

Page 2: ...Toco Zero 4 Indicators and Connectors 5 Power On Indication 5 RS232 Output 5 Remote Patient Event Marker 5 Probe Connection 5 Operating Procedures 6 User Configuration 6 The Printer 8 Ultrasound Feta...

Page 3: ...mydoppler de A service manual containing circuit descriptions diagrams parts and spares lists is available for the PFM1 by contacting the address above To maintain the PFM1 s performance it is recomm...

Page 4: ...ided with the system are connected to a properly grounded receptacle l Do not remove the system cover The system should be serviced by trained and qualified personnel only Contacting the hazardous vol...

Page 5: ...ansducer or cable damage could occur Symbols Used The following symbols are used on the PFM1 and are in accordance with BS EN60601 1 1990 Where they are associated with the connection of external equi...

Page 6: ...change user options Volume Down This button is used to decrease the audio volume level or in configuration mode to change user operations US1 US1 Transducer Select and Twins trace mode Pressing this...

Page 7: ...t can be applied to this output is 15VDC WARNING Any external equipment connected to this output must meet the relevant safety standard Remote Event Marker Connection of an external event mark switch...

Page 8: ...Up button One press advances the day by one pressing the Volume Up button for longer advances the day by 10 To change the year press the Volume Down button One press advances the year by one pressing...

Page 9: ...a Block is printed at the start of each recording this function can be disabled by setting Data Block to off It changes from on to off and back again by pressing the Volume Up button When the Data Blo...

Page 10: ...al side of the paper is uppermost Pull the paper out of the front of the unit making sure it is positioned below the door catches Close the door and latch by pressing down firmly on it To operate the...

Page 11: ...only showing an outline heart then this is an indication that the signal quality is not optimum Improved recordings can be obtained by repositioning the transducer so that this indicator always shows...

Page 12: ...the ingress of fluids NO coupling gel is required for this transducer to operate correctly Place the transducer on the centre line over the fundus in a position where the uterus is firm and secure in...

Page 13: ...he NST is summarised Reactive accelerations of 15 BPM or more which last for 15 seconds or longer twice in any 10 minute period These may be accompanied by fetal activity Non reactive evidence of feta...

Page 14: ...checking fuses and connections Check all connections and fuses For replacement fuses consult the technical specifications for specific ratings required No display information on LCD Verify the system...

Page 15: ...se on any damaged part contact them immediately Cleaning Enclosure Clean the exterior of the system with a soft dry cloth In the event of stubborn spots disconnect the system from the power source Use...

Page 16: ...Data Presentation Strip chart recorder and alphanumeric display module Printhead 4 inch thermal solid state printhead Resolution 8 dots mm Speeds 1 2 3 cm min Paper Roll or Z fold Paper type Plain bla...

Page 17: ...lving adverse EMC conditions Emissions Care has been taken through the design and manufacturing processes to minimise the EM emissions that may be produced by this equipment However in the unlikely ev...