SonixTouch Q+ Service Manual
00.053.204, Revision A
Chapter 4: Performance Testing
4-3
4.5
PULSED DOPPLER SOUND TEST
To Conduct a Pulsed Doppler Sound Test:
1.
Ensure the system is powered on and has completed the initialization sequence.
2.
Ensure the system has an exam screen showing.
3.
Tap the touch screen
PW
button.
4.
Press the console button.
5.
Tap
Audio
and turn the associated dial clock-wise until it reaches 100% .
6.
The sound of static should be audible.
4.6
LICENSES
Check system licenses to ensure at least some licenses are
Enabled
. If no licenses are
Enabled
, repeat section
System Date and Install Licenses
To Check System Licensing:
1.
Tap the touch screen
Menu
button.
2.
Select
Administrator > Licensing
.
3.
If none of the licenses are
Enabled
, repeat section
3.4 Set System Date and Install Licenses
or reload
license.key
following the instructions in section
Note:
There is a level indicator on the LCD display.
for more information on the touch screen dials.
Note:
If static is not audible, use the trackball to move the gate away from the top of the LCD display.
Summary of Contents for SonixTouch Q+
Page 1: ...Analogic Ultrasound Service Manual SonixTouch Q Ultrasound System...
Page 2: ......
Page 4: ......
Page 14: ...Chapter 2 System Specifications 00 053 204 Revision A SonixTouch Q Service Manual 2 2...
Page 42: ...Chapter 4 Performance Testing 00 053 204 Revision A SonixTouch Q Service Manual 4 4...
Page 120: ...Chapter 8 DICOM 00 053 204 Revision A SonixTouch Q Service Manual 8 24...
Page 147: ......