Page
16
of
20
Issue 9
– 17122020
Recommended separation distances between
portable and mobile RF communications equipment and the ME equipment
TheAlternating Pressure Air Flotation Mattress is intended for use in an electromagnetic environment in which radiated
RF disturbances are controlled. The customer or the user of theAlternating Pressure Air Flotation Mattress can help
prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications
equipment (transmitters) and the
Alternating Pressure Air Flotation Mattress
as
recommended below, according to the maximum output power of the communications equipment.
Rated maximum output
power of transmitter
W
Separation distance according to frequency of transmitter
m
150 kHz to 80 MHz
80 MHz to 800 MHz
800 MHz to 2.5 GHz
3,5
3,5
7
d=
P
d=
P
d=
P
1
1
1
0.01
0.12
0.12
0.23
0.1
0.37
0.37
0.74
1
1.17
1.17
2.33
10
3.7
3.7
7.37
100
11.67
11.67
23.33
15 EMC Information