anufacturer’s declaration-electromagnetic immunity
The device(s) is intended for use in the electromagnetic environment (for home and professional healthcare) specified below.
The customer or the user of the device(s) should assure that it is used in such an environment.
Immunity test
IEC 60601
test level
Compliance level
Electromagnetic environment-
guidance (for home and
professional healthcare environment)
Electrostatic
discharge(ESD)
IEC 61000-4-2
Contact:
±
8 kV
Air
±
2 kV,
±
4 kV,
±
8 kV,
±
15 kV
Contact:
±
8 kV
Air
±
2 kV,
±
4 kV,
±
8 kV,
±
15 kV
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%
Electrical fast
transient/burst
IEC 61000-4-4
+ 2kV for power supply lines
+ 1kV for input/output lines
+ 2kV for power supply lines
Not applicable
Mains power quality should be that of a
typical home and professional
healthcare environment.
Surge
IEC 61000-4-5
+ 0.5kV, +1kV line(s) to line(s)
+ 0.5kV, +1kV,+ 2kV line(s) to
earth
+ 0.5kV, +1kV line(s) to line(s)
Not applicable
Mains power quality should be that of a
typical home and professional
healthcare environment.
Voltage Dips, short
interruptions and voltage
variations on power supply
input lines
IEC 61000-4-11
Voltage dips:
0 %
U
T
; 0,5 cycle
0 %
U
T
; 1 cycle
70 %
U
T
; 25/30 cycles
Voltage interruptions:
0 %
U
T
; 250/300 cycle
Voltage dips:
0 %
U
T
; 0,5 cycle
0 %
U
T
; 1 cycle
70 %
U
T
; 25 cycles
Voltage interruptions:
0 %
U
T
; 250 cycle
Mains power quality should be that of a
typical home and professional
healthcare environment. If the user of
the device(s) requires continued
operation during power mains
interruptions, it is recommended that the
device(s) be powered from an
uninterruptible power supply or a battery.
Power frequency(50, 60
Hz) magnetic field
IEC 61000-4-8
30 A/m
50 Hz or 60 Hz
30 A/m
50 Hz
The device(s) power frequency magnetic
fields should be at levels characteristic
of a typical location in a typical home
and professional healthcare
environment.
NOTE UT is the a.c. mains voltage prior to application of the test level.