10.  Transport Mode 

 

 

 
 
 
 

 

If the patient is being moved on the mattress, or there is a 

power cut, general pressure can be maintained in the system 

for an adequate period of time whilst disconnected from the 

mains. 

 

Simply disconnect air pipe connector and place the connector 

cover over it. The air pressures in the mattress will now 

equalise, but maintain a degree of comfort. This will maintain 

the cells in their present state for approximately 48 hours. 

 

It is important to restore the Heritage II Digital dynamic 

replacement mattress as quickly as possible by reconnecting 

the supply tubes to the power unit. 

 

11.  Alarms & Fault Findings 

 

The Heritage II Digital is equipped with audible and visual alarm indicators. These alert the user to the 

status of the available mains supply and any mattress defect. 
 

11.1  Low pressure alarm 

 

Upon detection of low pressure, an audible alarm will be heard and the 

Low Pressure

 indicator will 

illuminate. The audible alarm may be cancelled by pressing the 

Alarm Mute

 button.

 

The 

Low 

Pressure

 indicator will continuously illuminate until the low pressure fault condition is resolved.  

 

This condition could be caused, for example, by incorrect fitting of the air pipe connector, opening of 

the CPR valve or a leak in the mattress due to a cut or puncture. 
 

11.2  Power failure alarm 

 

If at any time the mains power should be removed from the power unit or the power cable is 

unplugged without turning the power unit OFF, an audible alarm will be heard and the 

Power Failure

 

indicator will illuminate. The audible alarm may be cancelled by pressing the 

Alarm Mute

 button. 

 

 

 

Note: When the power unit has not been used for more than 3 months, it might 
require 60 minutes operating time (or more) for the Alarm to function correctly.

 

 

11.3  Service (alternating failure alarm) 

 

Should your system develop a fault condition whilst in use, an audible alarm will be heard and the 

Service

 indicator will illuminate. Please refer to 

Table 1

 for error codes and call Ultimate Healthcare 

or your local distributor.

  

 
 

Summary of Contents for UPRA3678DT-2

Page 1: ...Heritage II Digital Turn Dynamic Mattress Replacement System UPRA3678DT 2 User Manual Issue 5 16 12 2020 UPRA3678DT 2 UM 5...

Page 2: ...2 Comfort level 12 8 3 Auto detection 12 8 4 Control panel lockout 12 8 5 Function mode switch 13 8 5 1 Alternate 13 8 5 2 Static 13 8 5 3 Constant low pressure 13 8 5 4 Turning feature 13 8 6 Auto F...

Page 3: ...Page 3 of 28 Issue 5 16 12 2020 15 Service and Maintenance 20 15 1 General 20 15 2 Fuse replacement 20 15 3 Air filter replacement 20 16 EMC Information 21...

Page 4: ...y slightly with respect to the basic design of the product 2 Intended Use The intended use of this product is to prevent and or manage pressure ulcers while optimising patient comfort for patients up...

Page 5: ...itate Mattress Replacement Top of the Mattress Bottom of the Mattress CPR valve Cable Management Fixings Air Pipe Connector 3 x Static Head Cells 11 x Standard Cells Removable Cover 4 x Smaller Lower...

Page 6: ...order to prevent personal injury BF symbol indicates this product is according to degree of protection against electric shock for type BF equipment The operator must read this document User Manual bef...

Page 7: ...pplied product Ensure the power cable is not trapped or twisted and is routed suitably to avoid crushing or entrapment when connected to the product Do not use your mattress system power unit in the p...

Page 8: ...ment pressure 70 kPa 101 3 kPa Water dust protection classification IP21 6 2 Mattress Dimensions 2000mm x 890mm x 200mm L x W X H Weight 7 4 kg No of cells 19 Cover material 2 way stretch polyester wi...

Page 9: ...ed Check the Yellow Turn Valves and make sure they are positioned in the non turn mode There are security straps fitted to the base cover of the mattress which should be fastened loosely to convenient...

Page 10: ...tor to the power unit and then ensure that the air tubes are free from any kind of obstruction and are not kinked Plug the power cable into a suitable electrical socket and switch ON using mains power...

Page 11: ...mfort Control Buttons 7 Operate Standby Button Indicators 3 Comfort Setting Indicators 8 Cycle Time Selection Button Alternate 4 Control Panel Lockout Button Indicator 9 Cycle Time Display 5 Function...

Page 12: ...o Detection Once activated the indicator will flash and the system will start detecting and will set the internal pressure for the appropriately The indicator light will flash until the patient s weig...

Page 13: ...he cells are not alternating and the internal cell pressure within the mattress is set at 15mmhg to provide a comfortable static surface 8 5 4 Turning feature Alternating Mode only Turning the Yellow...

Page 14: ...ssing the Cycle button Selections can be made from 10 30 minutes at 5 minute intervals The cycle time will be displayed in the LCD screen 8 8 Alarm mute The Alarm Mute button temporarily resets the au...

Page 15: ...he Low Pressure indicator will illuminate The audible alarm may be cancelled by pressing the Alarm Mute button The Low Pressure indicator will continuously illuminate until the low pressure fault cond...

Page 16: ...Auto Firm Completion Auto Firm Ended 6 Static ONCE Operate Static Completion Static Ended 7 N A Power Failure REPEAT Cycle 4 sec Power Off Power Failure Alarm No Display 8 Low Pressure REPEAT Cycle 4...

Page 17: ...Check the power unit is connected to the mains power supply and that the mains switch is turned ON The power unit is operating but the mattress is not alternating Ensure that the mattress inflation pr...

Page 18: ...nit in water Caution When cleaning disinfecting ensure that only a damp cloth is used Caution After cleaning dry the mattress out of direct exposure to sunlight Caution Using inappropriate detergents...

Page 19: ...le dried or air dried They may be tumble dried on a low heat for up to 90 minutes Drying temperature must not exceed 400C The mattress cover and mattress must be dry prior to refitting Caution Exceedi...

Page 20: ...r signs of wear or damage Ensure mattress cover and tubes are stubbed together correctly 3 Check the air hoses for any kink or break For replacement please contact Ultimate Healthcare or your local di...

Page 21: ...d professional healthcare environment RF emissions CISPR 11 Group 1 The device s uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any i...

Page 22: ...nvironment Surge IEC 61000 4 5 0 5kV 1kV line s to line s 0 5kV 1kV 2kV line s to earth 0 5kV 1kV line s to line s Not applicable Mains power quality should be that of a typical home and professional...

Page 23: ...ands between 0 15 MHz and 80 MHz 80 AM at 1 kHz 10 V m 80 MHz 2 7 GHz 80 AM at 1 kHz Portable and mobile RF communications equipment should be used no closer to any part of the device s including cabl...

Page 24: ...t power of transmitter W Separation distance according to frequency of transmitter m 150 kHz to 80 MHz d 1 2 P 80 MHz to 800 MHz d 1 2 P 800 MHz to 2 7 GHz d 2 3 P 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0...

Page 25: ...0 2 0 3 9 9 745 780 810 800 960 GSM 800 900 TETRA 800 iDEN 820 CDMA 850 LTE Band 5 Pulse modulation b 18 Hz 2 0 3 28 28 870 930 1 720 1 700 1 990 GSM 1800 CDMA 1900 GSM 1900 DECT LTE Band 1 3 4 25 UMT...

Page 26: ...Page 26 of 28 Issue 5 16 12 2020...

Page 27: ...Page 27 of 28 Issue 5 16 12 2020...

Page 28: ...ue 5 16 12 2020 Ultimate Healthcare Ltd Calmore Industrial Estate Nutwood Way Totton Southampton Hampshire SO40 3WW Tel 0333 321 8996 Fax 023 8066 2388 Email info ultimatehealthcare co uk www ultimate...

Reviews: