Instructions for use (rev 050320)
PRE
UGIN DENTAIRE
16
3.2.
CONDITIONS OF USE, TRANSPORT AND STORAGE
Authorised conditions of use
-
Ambient temperature: +5°C to +40°C
-
Humidity range: maximum relative humidity 80% for temperatures until 31°C, without condensation and
decrease linear until 50% to 40°C, without condensation
-
Ambient pressure: up to an altitude of 2000 m above sea level
Authorised conditions of storage
-
Ambient temperature: -20°C to +65°C
-
Humidity range: maximum relative humidity 80%
-
Ambient pressure: up to an altitude of 2000 m above sea level
Conditions to transport
-
Wait the complete cooling before packing/transporting
-
Use the original packaging only, with the protective elements
-
Install the equipment in the packaging with the protective elements.
4.
INSTALLATION AND START-UP
4.1.
UNPACK
-
Unpack PRE and check that it is in perfect order
-
Any defects can be reported to the carrier
-
Remember to get the carrier to sign the delivery note
NOTE: keep the original packaging to transport the equipment (see chapter 3.2)
4.2.
INSTALLATION
-
Install in a room aired. Observe the conditions indicated in chapter 3.2
-
Place the equipment on a level and aired surface,
easily accessible
. Keep a distance from other objects
around not less than 10 cm.
-
Keep it away from heat sources (for example, radiators and/or other equipment that release heat).
Avoid vibrations and shocks
-
Do not install in a place with explosion hazard
-
Its forbidden placing inflammables, toxic, volatile or explosives around the equipment
Grounding is mandatory to avoid accidents and the dangers of overheating the power line.
Due to the power of equipment: connect it directly to the mains, do not use multiple sockets
(risk of electrical interference)
Summary of Contents for PRESSOJET +