TEMPERATURE CALIBRATION
The furnace have been subjected to a temperature test using a temperature calibrator. The
setting is exact (SET VALUE = 5) during factory calibration and while the muffle is new. However,
should you not find the initial setting satisfactory; you can change the set value.
8.1.
Temperature calibration change
To change this setting:
-
press both the
and
keys at the same time.
-
Go to page 2 : press PAGE key
-
The display screen will show:
-
If the result indicates :
o
on overly high firing temperature: increase the setting value (6 or 7)
o
an inadequate firing temperature: decrease the setting value (4 or 3)
Set value
Value
Key
7
= 2 points = -10°C
and
to confirm
6
= 1 point = -5°C
5
Factory calibration
4
= 1 point = +5°C
and
to confirm
3
= 2 points = +10°C