8.3 Overview of solutions to possible malfunctions
The table below lists possible component malfunctions and the associated solutions. This will help you to
save time and the cost of a visit from a technician. If it is necessary to change any component, please use
only original spare parts. Those types of repairs should only be performed by a trained technician. For any
type or repair, the surgical chair must be disconnected from the power outlet.
Malfunction
Component
Reason
Solution
Power source is
missing
Plug in the power cord
Power cord
Failure of the power
cord
Replace the power cord
Cable of the control is
not fully plugged in
Hand control/Foot control
Plug cable into the
control
None of the motors
will run
Hand control defect
Replace hand control
Fuses
Power supply fuse is
blown
Replace the
corresponding fuse
Replace the battery
and or adapter
Battery
Defective battery
Control box
Control box defect
Replace the control box
Motor connection cables
Motor connection
cables not plugged in
Plug in motor
connection cables
One of the motors
is not running
Defect in motor
connection cable
Replace motor
connection cable
Motor
Motor defect
Replace the entire
motor
Corresponding button
is defective
Hand control/Foot control
Replace the control
Control box
Control box defect
Replace the control box