Resetting the Thermometer

If you want to retore the thermometer settings as delivered from the 

factory, press “       “ or  “       “ until the display shows “       “  then 

press “       “ to display “       “ in the primary display. Press “       “ to 

exit setup.

2. How to Measure Temperature

Connecting a Thermocouple

Thermocouples are color coded by based on the ANSI color code:

    1.  Plug a thermocouple into the input terminal(s).

    2.  Set the thermometer for the correct thermocouple type. To  

      change the thermocouple type, see “

How to Change  

      Setup Options

”.

Displaying Temperature

    1.  Press “       “ for more than 1 second to select the correct  

      temperature unit.

    2.  Hold or attach the thermocouple(s) to the measurement location.  

      The temperature reading appears in the selected display. 

    3.  (DT302)   Press “       “ momentarily to show T1 reading in the  

       

primary display, and toggle showing T2 reading, or  

       

the internal temperature of the thermometer in the  

       

secondary display.

       

Press “       “ for more than 1 second to show T1-T2  

       

reading in the primary display.

       

Press “       “ momentarily to toggle showing T1, T2,  

       

or the internal temperature of the thermometer in the  

       

secondary display while the primary display shows  

       

the T1-T2 reading.

   NOTE:

 

The display shows “

----

” when a thermocouple is not  

      connected. The display shows “      “ (OVERLOAD) when the  

      temperature being measured is outside the thermocouple’s  

      valid range.

9

Type  Color Type  Color

   J 

Black 

   T 

Blue

   K 

Yellow 

   E 

Purple

    4.  If you selected a user-defined logging interval (1 second to  

      23 hours and 59 minutes).

      •  Press “       “ or  “       “ until the display shows “

h:m

” or  

         “

m:s

”, and then press “       “ to select. The last number in the  

         time display blinks.

      •  Press “       “ or  “       “ until the logging interval you want  

         appears on the display, and then press “       “ to select.

      Holding down “       “ or  “       “ causes the number to change  

      more quickly.

Changing the Offset

You can adjust the thermometer’s readings to compensate for the 

errors of a specific thermocouple. The allowable compensation range 

is ±5.0˚C or K, and ±9.0˚F. (DT302 you can store individual offsets for 

T1 and T2)..

    1.  Press “       “ or “       “ until the display shows “ 

OFFSET 

”.

    2.  Press “       “ to display T1 reading (blinks), T1 offset.

    3.  (DT302)   Press “       “ to display T1 reading (blinks), T1 offset.

       

Press“       “ or “       “ to display T1 or T2 in the

       

primary display to select 1 input.

      (DT301)   Press “       “ to display reading and offset.

       

The temperature measurement plus the offset  

       

appears in the primary display. The offset appears in  

       

the secondary display.

    3.  Press “       “ or “       “ until the primary display shows the  

      correct reading, and then press “       “ to store the offset setting  

      in memory.

      Remember to reset the offset to 0.0 when it is no longer needed.

   NOTE:

 

If T1 or T2 has no offset at all, the display does not  

      show “ 

OFFSET

 ”.

8

www.

GlobalTestSupply

.com

Find Quality Products Online at:

[email protected]

Summary of Contents for DT301

Page 1: ...DT301 DT302 Digital Temperature Logger INSTRUCTION MANUAL www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...

Page 2: ...the proper thermocouple type you are using Be sure the thermocouple you use can withstand the temperature extreme it may be exposed to in your service task Properly maintain your thermometer and cali...

Page 3: ...C F K Temperature units 11 Secondary Display Internal temperature of the thermometer MEMORY MIN MAX AVG and T1 or T2 DT302 reading Shows the memory location initially when a logged reading is recalled...

Page 4: ...ss to accept the selected setup option 5 Buttons NOTE Press all the dual function buttons momentarily to activate the Upper Row Function and press these buttons for more than 1 second to activate the...

Page 5: ...display shows and LOG 2 Press to display the logging interval choices 3 Press or until the display shows the logging interval you want and then press to select 7 Setting the Date 1 Entering Setup will...

Page 6: ...erature being measured is outside the thermocouple s valid range 9 Type Color Type Color J Black T Blue K Yellow E Purple 4 If you selected a user defined logging interval 1 second to 23 hours and 59...

Page 7: ...set of initial conditions when logging continuously It stores 9 999 sets of temperature readings and initial conditions when logging individual points manually 11 Holding the Displayed Temperature 1...

Page 8: ...Starting and Stopping Logging Memory clear and PC communications are in accessible during logging Recall function is enabled during logging 1 Set the time and the logging interval see How to Change Se...

Page 9: ...ery 4 Reattach the battery compartment cover to the meter and reinstall the screws 15 These guidelines will help you attain long and reliable service from your meter 1 Calibrate your meter annually to...

Page 10: ...cified for ambient temperatures between 64 F 18 C and 82 F 28 C for a period of 1 year The above specifications do not include thermocouple error 16 Specifications Environmental General Thermocouple O...

Reviews: