Insert the microphone into the hole of the calibrator. Press calibrator        

key to power on, and adjust the CAL potentiometer of the unit, the

level display will indicate the desired level. While selecting C weighting,
d/-0.3 dB deviation is reasonable.

Replacing the Battery

When the icon of 

"         "

appears, the 9V battery has fallen to 

a critically low voltage level and should be replaced as soon as 
possible. Use a screwdriver to unscrew the back battery compartment
cover. Insert 4 fresh AAA batteries and screw the cover back on.

Diagram O

1.

Calibration Adjustment

(side) By using a Phillips (+)
screw driver to adjust the 
calibration (Follow the 
calibration instruction).

2.  

Tripod Mount

Mount the meter to a 
camera tripod for increased 
stability and accuracy, further
eliminate hand or any sound
reflected form the user.

CE Certification

The meter conforms to the following standards:

EN 50081-1/1992 : EN 55022
EN 50082-1/1997 :
(EN 61000-4-2/-3/-8, ENV 50204)

The meter complies with the essential protection requirements of
Council Directive 89/336/EEC on the approximation of the laws of the
Member States relating to electromagnetic compatibility.

If the meter is operating in manual range and 

"LO"

is displayed, the

sound is too low or the range.

If 

"HI"

is displayed, the sound is too loud. In either case, you must

adjust the measurement range or your readings will be inaccurate.

Diagram L

Diagram M

Automatic Shutoff

The meter will turn off automatically after 20 minutes to preserve 
the battery.

To override this feature:

1.  Make sure the unit is turned off.

2.  Press the      and 

"MAX HOLD"

buttons simultaneously.

3.  When the full display appears, release the        button, 

"       "

icon will appear on the screen, now release the 

"MAX

HOLD"

button, 

"      "

will be replaced with full display.

4.  The meter will remain on until the        button is pressed again.

The automatic shut off feature will resume the next time the
meter is turned on.

Diagram N

Calibration

Using a standard Acoustic Calibrator which generate "94 dB" output.

1.  Set whichever time weighting: Fast or Slow.

2.  Set sound level range at: 80-110 dB.

3.  Select weighting A or C.

4.  MAX HOLD function measurement mode disabled.

5.  Calibrate under 60 dB sound environment. (recommended)

DSM101-MAN    

P. 3

Summary of Contents for DSM101

Page 1: ...1 800 547 5740 Fax 503 643 6322 www ueitest com email info ueitest com DSM101 INSTRUCTION MANUAL ...

Page 2: ... the meter on The unit will first display the full screen and 188 8 then count down from 99 9 to zero The meter will now begin measuring the current sound levels SPL Sound Pressure Levels appears on the left side A dB on the right side of screen Point the microphone toward the source of the sound to be measured Diagram A Diagram B Selecting the Response Time You can select fast or slow response ti...

Page 3: ...t the two digit numbers to the left of the bar graph will change to reflect the low of the newly selected range 2 Press RNG key to switch back to automatic ranging C weighting is suitable for the flat response measurements with no increase or decrease of amplitude over the frequency spectrum Applications for C weighting include the sound level analysis of engines and machinery Press the A C key to...

Page 4: ...irements of Council Directive 89 336 EEC on the approximation of the laws of the Member States relating to electromagnetic compatibility If the meter is operating in manual range and LO is displayed the sound is too low or the range If HI is displayed the sound is too loud In either case you must adjust the measurement range or your readings will be inaccurate Diagram L Diagram M Automatic Shutoff...

Page 5: ...ited to implied warranties of merchantability and fitness for a particular purpose are limited to the express warranty UEi shall not be liable for loss of use of the instrument or other incidental or consequential damages expenses or economic loss or for any claim or claims for such damage expenses or economic loss A purchase receipt or other proof of original purchase date will be required before...

Reviews: