4. Press the 

"REC"

key again. "

MAX"

will appear on screen side 

by the 

"REC"

and the maximum sound level recording at this

time, but the bar graph will continue to show the current reading.

Diagram I

5. Press the 

"REC"

key again to resume recording and repeat 

the process.

6. Press and hold the 

"REC"

key until the 

"REC"

indicator 

disappears to exit recording mode.

Selecting Automatic and 

Manual Ranging

The meter features 4 measurement ranges in 10 dB steps: 40-70 dB. 
60-90 dB, 80 dB - 110 dB, 100 dB - 130 dB.

When you turn the meter on, it will be in automatic range mode and a
small "

AUTO

" will be displayed on the left side of the screen. In this

mode, the meter will adjust the measurement range automatically 
for accuracy.

The two digit number to the left of the bar graph on the LCD will show
the low end of the current range. You can also set the range manually.

Diagram J

Diagram K

In the manual range mode, the meter will be able to take readings more
quickly, because the unit does not need to first establish the range
before displaying the measurement. This is helpful when you know the
measurement range in advance.

To adjust the range manually:

1. When measuring sound levels, press the 

"RNG"

keys as needed 

to adjust the measurement range. 

MANU

will appear on the 

display. Note that the two digit numbers to the left of the bar
graph will change to reflect the low of the newly selected range.

2. Press "

RNG

" key to switch back to automatic ranging.

C weighting is suitable for the flat response measurements with 
no increase or decrease of amplitude over the frequency spectrum.
Applications for C weighting include the sound level analysis of engines
and machinery.

Press the 

"A/C"

key to select between A and C weighting. A small 

"A"

or 

"C"

icon will be displayed on the right side of the screen to indicate

the current mode.

Freezing the Maximum Sound 

Level Reading

1.  Press the        key to turn the meter on.

2.  When measuring sound levels, press the 

"MAX HOLD"

key 

to freeze the maximum reading. 

"MAX HOLD"

will be 

displayed. The digital display will remain unchanged until 
a higher reading is detected. Note that the bar graph will 
continue to record the current reading.

3.  Press the 

"MAX HOLD"

key again to exit maximum 

hold mode.

Diagram F

Recording the Maximum and 

Minimum Measurements

1.  Press the         key to turn the meter on.

2.  Press the 

"REC"

key. The 

"REC"

icon will be displayed in 

the upper corner of the screen. The meter will begin tracking 
the maximum and minimum sound level measurements.

3.  Press the 

"REC"

key again. 

"MIN"

icon will appear in the

upper middle of the screen and the minimum sound level 
measurement will be displayed. The unit is not recording 
at this time, but the bar graph will continue to show the 
current reading.

Diagram G

Diagram H

DSM101-MAN    

P. 2

Summary of Contents for DSM101

Page 1: ...1 800 547 5740 Fax 503 643 6322 www ueitest com email info ueitest com DSM101 INSTRUCTION MANUAL ...

Page 2: ... the meter on The unit will first display the full screen and 188 8 then count down from 99 9 to zero The meter will now begin measuring the current sound levels SPL Sound Pressure Levels appears on the left side A dB on the right side of screen Point the microphone toward the source of the sound to be measured Diagram A Diagram B Selecting the Response Time You can select fast or slow response ti...

Page 3: ...t the two digit numbers to the left of the bar graph will change to reflect the low of the newly selected range 2 Press RNG key to switch back to automatic ranging C weighting is suitable for the flat response measurements with no increase or decrease of amplitude over the frequency spectrum Applications for C weighting include the sound level analysis of engines and machinery Press the A C key to...

Page 4: ...irements of Council Directive 89 336 EEC on the approximation of the laws of the Member States relating to electromagnetic compatibility If the meter is operating in manual range and LO is displayed the sound is too low or the range If HI is displayed the sound is too loud In either case you must adjust the measurement range or your readings will be inaccurate Diagram L Diagram M Automatic Shutoff...

Page 5: ...ited to implied warranties of merchantability and fitness for a particular purpose are limited to the express warranty UEi shall not be liable for loss of use of the instrument or other incidental or consequential damages expenses or economic loss or for any claim or claims for such damage expenses or economic loss A purchase receipt or other proof of original purchase date will be required before...

Reviews: